![Stryker 6516 Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/stryker/6516/6516_operation-manual_1385843054.webp)
C
Clle
ea
an
niin
ng
g tth
he
e cch
ha
arrg
ge
err
W
WA
AR
RN
NIIN
NG
G
•
Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the S
SM
MR
RT
T Pak to
reduce the risk of injury.
•
Always disconnect the S
SM
MR
RT
T charger from the wall outlet before cleaning to avoid the risk of electrical hazards.
•
Do not spray fluid directly onto the S
SM
MR
RT
T charger.
•
Do not power wash the S
SM
MR
RT
T charger.
•
Do not use solvents, lubricants, or other chemicals to clean the S
SM
MR
RT
T charger unless otherwise directed.
•
Do not immerse the S
SM
MR
RT
T charger in water or allow water to collect on top of the S
SM
MR
RT
T charger to avoid the risk of
electric shock.
To clean the S
SM
MR
RT
T charger:
1. Disconnect the S
SM
MR
RT
T charger from the wall outlet to avoid electrical hazards during cleaning.
2. Wipe surfaces of the S
SM
MR
RT
T charger with a soft cloth dampened with a non-abrasive, hospital disinfectant.
3. Wipe with a cloth moistened with clean water to remove any cleaning chemicals or residue.
4. Dry thoroughly before placing the S
SM
MR
RT
T charger back into service.
C
Clle
ea
an
niin
ng
g tth
he
e b
ba
atttte
erryy
The S
SM
MR
RT
T Pak is designed to be power washable. The preferred method for power washing the S
SM
MR
RT
T Pak is with the
standard hospital surgical cart washer or hand held wand unit.
W
WA
AR
RN
NIIN
NG
G
•
Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power
washing equipment may aerate contamination.
•
Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the S
SM
MR
RT
T Pak to
reduce the risk of injury.
•
Always use only non-conductive materials to wipe the S
SM
MR
RT
T Pak.
•
Always avoid excessive water exposure to the S
SM
MR
RT
T Pak terminals.
•
Always refer to the disinfectant’s Material Safety Data Sheet (MSDS) to verify the pH range. Disinfectants with pH levels
higher than 10.5 may cause the S
SM
MR
RT
T Pak housing material to crack.
•
Do not directly handle or make contact with the S
SM
MR
RT
T Pak terminals while cleaning to avoid the risk of injury.
•
Do not immerse the S
SM
MR
RT
T Pak in liquid to reduce the risk of electric shock.
•
Do not use solvents, lubricants, or other chemicals to clean the S
SM
MR
RT
T Pak unless otherwise directed.
C
CA
AU
UT
TIIO
ON
N
•
Do not steam clean or ultrasonically clean the S
SM
MR
RT
T Pak.
•
Do not exceed 180 °F/82 °C as the maximum water temperature.
•
Do not exceed 240 °F/115 °C as the maximum air dry temperature (cart washers).
•
Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you use a hand held wand to wash the product,
keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
To clean the S
SM
MR
RT
T Pak:
1. Remove the S
SM
MR
RT
T Pak from the P
Po
ow
we
err--P
PR
RO
O cot or S
SM
MR
RT
T charger.
2. Inspect the S
SM
MR
RT
T Pak housing and terminal area for any cracks or damage.
3. Clean the S
SM
MR
RT
T Pak with a hospital grade disinfectant with a pH range of 6.5 to 10.5.
4. Rinse the S
SM
MR
RT
T Pak thoroughly with clean water to remove any cleaning chemical or residue. Position the S
SM
MR
RT
T Pak
to avoid water from pooling near the terminals.
5. Dry the S
SM
MR
RT
T Pak thoroughly before insert the S
SM
MR
RT
T Pak into a P
Po
ow
we
err--P
PR
RO
O cot or S
SM
MR
RT
T charger.
EN
50
6516-209-001 Rev D.0
Содержание 6516
Страница 2: ......
Страница 62: ...6 65 50 06 60 00 00 01 10 00 01 16 6 Rev AA A A 650600010130 2 EN 58 6516 209 001 Rev D 0 ...
Страница 63: ...6 65 50 00 0 0 00 02 2 0 02 28 8 Rev A A A 6500 001 160 1 6516 209 001 Rev D 0 59 EN ...
Страница 64: ...6 65 50 00 0 1 10 02 2 0 01 15 5 Rev AA A A 6500 102 014 1 EN 60 6516 209 001 Rev D 0 ...
Страница 66: ...6 65 50 00 0 0 00 02 2 1 15 59 9 Rev AA A A 6500 002 159 1 EN 62 6516 209 001 Rev D 0 ...
Страница 67: ...6 65 50 00 0 2 20 01 1 1 14 48 8 Rev AA A A 6500 201 248 1 6516 209 001 Rev D 0 63 EN ...
Страница 68: ...6 65 50 00 0 2 20 01 1 0 01 10 0 Rev AA A A 6500 201 140 1 EN 64 6516 209 001 Rev D 0 ...
Страница 70: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6516 209 001 Rev D 0 WCR AA 5 2019 12 ...