F
Fiig
gu
urre
e 2
25
5 –– T
Th
hiig
gh
h rre
essttrra
aiin
ntt llo
occa
attiio
on
n
A
Atttta
acch
hiin
ng
g tth
he
e a
an
nkklle
e rre
essttrra
aiin
nttss
To attach the X-restraint ankle restraints:
1. Wrap the restraint around the cot frame where the matching label is located (Figure 26).
2. Pull the restraint buckle through the loop, toward the head end of the cot.
3. Pull the buckle tight.
4. Connect the patient right buckle to the patient left buckle.
F
Fiig
gu
urre
e 2
26
6 –– A
An
nkklle
e rre
essttrra
aiin
ntt llo
occa
attiio
on
n
A
Ad
djju
ussttiin
ng
g rre
essttrra
aiin
ntt ssttrra
ap
pss
Open the restraint straps and place them at either side of the cot while you position the patient on the cot mattress.
Lengthen the restraint straps, buckle them around the patient, and shorten them to tighten.
•
To open the restraint strap, press the red button on the front of the buckle receiver. This allows you to release the buckle
latch plate and pull it out of the receiver.
•
To close the restraint strap, push the latch plate into the receiver until you hear a click.
•
To lengthen the restraint strap, grasp the buckle latch plate, turn it at an angle to the webbing, then pull it out. A hemmed
tab at the end of the webbing prevents the latch plate from coming off of the strap.
•
To shorten the restraint strap, grasp the hemmed tab and pull the webbing back through the latch plate to tighten.
When you buckle a restraint strap around a patient, secure the latch plate and remove any loose webbing from the cot.
Inspect the restraint straps at least once a month (more if used often). Check for a bent or broken receiver or latch plate, or
torn or frayed webbing. Replace any worn or inoperable restraint strap.