6. Reverse steps to reinstall.
7. Verify proper operation before you return the product to service.
M
Ma
an
nu
ua
all rre
elle
ea
asse
e b
bu
utttto
on
n a
asssse
em
mb
bllyy rre
em
mo
ovva
all a
an
nd
d rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T25 Torx drive
•
5/32'' hex wrench
P
Prro
occe
ed
du
urre
e::
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
2. Remove the trolley covers. See
3. Press the main power button to turn the product off.
4. Remove the slic pin (A) from the switch/bracket assembly (Figure 15).
5. Using a T25 Torx driver, remove the two button head cap screws (B) that secure the manual release button bracket to
the trolley frame (Figure 15).
6. Using a 5/32" hex wrench, loosen (do not remove) the socket head cap screw (C) that secures the manual release
button bracket to the trolley frame (Figure 15).
6390-309-002 Rev AB.0
37
EN
Содержание 6390
Страница 2: ......
Страница 6: ......
Страница 77: ...A An nc ch ho or r a as ss se em mb blly y 6390 001 011 Rev AC Reference only 6390 309 002 Rev AB 0 71 EN ...
Страница 85: ...6390 309 002 Rev AB 0 79 EN ...
Страница 91: ...6390 309 002 Rev AB 0 85 EN ...
Страница 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Страница 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Страница 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Страница 98: ...EN 92 6390 309 002 Rev AB 0 ...
Страница 99: ...6390 309 002 Rev AB 0 93 EN ...
Страница 100: ...EN 94 6390 309 002 Rev AB 0 ...
Страница 101: ...6390 309 002 Rev AB 0 95 EN ...
Страница 102: ...EN 96 6390 309 002 Rev AB 0 ...
Страница 103: ...6390 309 002 Rev AB 0 97 EN ...
Страница 137: ......
Страница 138: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6390 309 002 Rev AB 0 WCR AE 5 2021 02 ...