![Stryker 4208 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/stryker/4208/4208_instructions-for-use-manual_1386101014.webp)
13
Guidance and manufacturer’s declaration - electromagnetic immunity
The System 5 handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of
the System 5 handpiece should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
test level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
n/a
n/a
3 V/m
80 MHz to 2.5 GHz
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the System 5 handpiece, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d=1.67√
P
d=1.67√
P
80 MHz to 800 MHz
d=2.33√
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in meters (m)
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.