17
Cleaning
Model 1231/1731 stretchers are designed to be power–washable. The unit may show some signs of oxidation
or discoloration from continuous washing. However, no degradation of the stretcher’s performance charac-
teristics or functionality will occur due to power washing as long as the proper procedures are followed.
S
Follow the cleaning solution manufacturer’s dilution recommendations exactly.
S
Remove the mattress prior to washing the unit; do not wash the mattress with the stretcher.
S
Position the Fowler at 45
_
, place the unit in full reverse Trendelenburg (foot end down), raise the siderails,
and place the I.V. poles and push handles in the up position.
S
Stryker Medical recommends the standard hospital surgical cart washer for power washing Model
1231/1731 stretchers.
S
Do not replace the mattress on the stretcher until the unit is completely dry.
DO NOT STEAM CLEAN THE UNIT. Use a maximum water temperature of 180
_
F/68
_
C. Maximum air dry
temperature (cart washers) is 240
_
F/115
_
C. Water pressure – 1500 psi/130.5 bar. If a hand held wand is
being used to wash the unit, the pressure nozzle must be kept a minimum of 24 inches/.61m from the unit.
Stretchers must have maintenance performed after a minimum of every fifth washing. Refer to the mainte-
nance manual for specific lubrication instructions.
Failure to comply with these instructions may invalidate any/all warranties.