Configuring the Wireless Settings for the SYNK Wireless Transmitter

The SYNK Wireless System uses a radio-frequency (RF) link to deliver high-definition video signals 
in real time� Each system uses a portion of the RF spectrum commonly referred to as a “channel” 
to broadcast its signal� Channels are commonly designated using the formula:

Channel Number = (Frequency in MHz – 5000) / 5

The SYNK Wireless System can use any available 20 MHz channel in the 5�15-5�825 GHz spectrum� 
The list of allowed channels is regulated and varies by region�

The 5�15-5�825 GHz spectrum is also used by 802�11a/n devices as well as 5�8 GHz cordless 
phones� The SYNK Wireless System uses an automatic frequency selection (AFS) mechanism in 
combination with a passive scanning capability to maintain compatibility with these devices� 
Using AFS and passive scanning, the system will identify an unoccupied 20 MHz channel for use� 
Each linked SYNK Wireless System requires one 20 MHz channel� Systems that are separated by a 
large distance may use the same channel�

Note: The SYNK wireless system uses a proprietary transmission method, which cannot 
connect to or be accessed by 802.11 devices

In order to make sure that there are enough free channels for each SYNK System, follow the steps 
outlined below:
1�  Contact your IT administrator to determine if 802�11a or 802�11n WiFi is in use at your facility� 

If so, note which channels are in use in or near each SYNK System installation location�

2�  Determine if any 5�8 GHz cordless phones are in use at your facility� If so, note their locations�
3�  For each SYNK System installation location, ensure that there are two unused channels per 

SYNK Transmitter installed� If a 5�8 GHz cordless phone is used in the vicinity, assume that 
channels 149-165 are occupied�

Contact your Stryker sales representative for more details�

Содержание 0240-031-000

Страница 1: ...SYNK Wireless System Transmitter 0240 031 010 Display 0240 031 000 ...

Страница 2: ......

Страница 3: ...Transmitter Front Panel 8 SYNK Wireless Transmitter Rear Panel 9 Setup and Interconnection 10 Permanently Linking the Transmitter to an Auxiliary Display 15 Removing the the Permanent Link 15 Troubleshooting 16 Cleaning and Maintenance 18 Cleaning the Display and Transmitter 18 Maintenance 19 Technical Specifications 20 Electromagnetic Compatibility 22 Symbol Definitions 26 ...

Страница 4: ......

Страница 5: ...or to a surgical procedure This system was fully tested at the factory before shipment Do not place any heavy object on the power cord Damage to the power cord can cause fire or electric shock This system is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Do not put any liquid or solid object into the display panel or the transmitter ...

Страница 6: ...rom a cold location to a warm location Pay close attention to the care and cleaning instructions in this manual A deviation may cause damage Do not install the devices near sunlight excessive dust mechanical vibration or shock Keep the devices away from equipment that uses strong magnets i e large loudspeakers Do not touch the patient with signal input or output connectors Equipment with SIP SOP c...

Страница 7: ...ician for help To ensure electromagnetic compatibility refer to the Electromagnetic Compatibility section of this manual The VisionPro SYNK 26 Wireless LED Display REF 0240 031 000 and the SYNK Wireless Transmitter 0240 031 010 must be installed and operated according to the EMC information provided in this manual The warranty is void if any of these warnings or cautions are disregarded These prod...

Страница 8: ...e display also supports a USB port and serial communication via the RS232 port Note For details and instructions on using the VisionPro SYNK 26 Wireless LED Display refer to the VisionPro User Manual P22286 This manual provides instructions for using the display and the transmitter together SYNK Wireless Transmitter REF 0240 031 010 The SYNK Wireless Transmitter allows the VisionPro SYNK 26 Wirele...

Страница 9: ...ral and gynecological surgery thoracoscopy endoscopy general gastroenterological and ENT surgery and general surgery The system is a non sterile reusable device not intended for use in the sterile field The system is intended for use by qualified physicians having complete knowledge of these surgical procedures Contraindications There are no contraindications for these products ...

Страница 10: ... 5 6 8 7 1 Full Screen mode button Shows one video feed on the display 2 Picture in picture mode button Shows a second video feed in the corner of the display in a smaller overlay 3 Picture by picture mode button Shows two video feeds side by side 4 Source Select button Toggles the video feeds shown in full screen picture by picture or picture in picture mode 5 Token Slot Token insertion site used...

Страница 11: ...esentative for more information 2 Digital Input Connects to a digital input source 3 Digital Output Connects to primary display 4 Audio Connectors Line in and Mic in connectors 5 Remotes Connectors Remote 1 and 2 connectors 6 RS232 Port Maintenance port 7 AC Power Inlet Connects to separable power cord that can be used for mains isolation 8 Fuse Panel Contains two 1 6A 250V slow blow high breaking...

Страница 12: ...10 Setup and Interconnection Connecting AC power Warning Always use an approved hospital grade power cord ...

Страница 13: ...at and shut down 1 Connect the DVI OUT 1 on the camera control unit to the DVI 1 IN on the SDC 2 Connect the DVI 1 OUT on the SDC to the DIGITAL INPUT on the transmitter 3 Connect the DIGITAL OUTPUT on the transmitter to the DVI on the primary display 4 Connect the C video input on the display to the C video output on the printer 5 Connect the C video input on the printer to the C video output on ...

Страница 14: ...12 IN OUT 4 1 2 5 3 6 Camera Control Unit SDC Transmitter Auxiliary Display Primary Display Printer ...

Страница 15: ...S and passive scanning the system will identify an unoccupied 20 MHz channel for use Each linked SYNK Wireless System requires one 20 MHz channel Systems that are separated by a large distance may use the same channel Note The SYNK wireless system uses a proprietary transmission method which cannot connect to or be accessed by 802 11 devices In order to make sure that there are enough free channel...

Страница 16: ... turns off insert the blue token into the token slot on the transmitter The token LED shines amber as it writes the data Stryker Endoscopy 5900 Optical Court San Jose CA 95138 USA 800 624 4422 Made in USA 0240031010 0 63A 250V RS232 AUDIO REMOTES 1 2 DIGITAL INPUT DIGITAL OUTPUT Place Serial Label Here EXPANSION CARD 100 240 50 60 Hz 630 mA FCC ID SSH SYNKTX This device complies with part 15 of th...

Страница 17: ...ay transmitter set has been permanently linked the set will automatically link when powered on Keep display transmitter sets that are in permanent link mode in the same operating room Removing the the Permanent Link To return the display s and transmitter to default linking mode please follow the instructions below using the token 1 Power on the display s and transmitter 2 Insert the token into th...

Страница 18: ...input is selected on the display and the transmitter Token LED blinking amber Please contact Stryer representative Abnormal picture Oversized undersized or missing display or center shift Using the Screen Menu adjust the Phase Frequency Horizontal and Vertical settings with non standard video signal timing Wait a few seconds after initial sync of video signals or power cycle the display OSD error ...

Страница 19: ...Current Status Remedy Wireless error symbol appears on the display Conditions are not sufficient for optimal wireless performance Switch to a wired video input on the display or contact your Stryker representative ...

Страница 20: ...s in any liquid as product damage will result Do not use corrosive cleaning solutions to clean the devices as product damage may result Do not sterilize the devices as product damage may result Cleaning the Display and Transmitter Display Note For complete display cleaning instructions refer to the VisionPro 26 LED Display manual P22286 Transmitter 1 Apply a standard disinfectant or mild detergent...

Страница 21: ...he latch 3 Replace the fuse with the same value and rating 4 Reinstall the fuse holder until the tab snaps in place Periodic Maintenance Schedule Warning To ensure safe operation of the transmitter you should periodically perform the following procedure Every 12 months check the earth leakage current to 500µA ground protective earth impedance to 0 1 ohms and power consumption less than or equal to...

Страница 22: ...cy 4 9 GHz 5 9 GHz Channel Bandwidth 20 MHz Channel Allocation Automatic frequency selection with prescan Protocol Orthogonal Frequency Division Multiplexing OFDM with Multiple Input Multiple Output MIMO Wireless Frequencies of operation 5 15 5 825 GHz Maximum Power Output 10 mW Channel Bandwidth 20 MHz Channel Allocation Automatic frequency selection with WiFi avoidance Encoding Orthogonal Freque...

Страница 23: ... respect to electric shock fire and mechanical hazards only in accordance with IEC 60601 1 and CAN CSA C22 2 No 60601 1 Continuous Operation Compliance FCC Regulations FCC 15B 2008 Class B FCC Identifier SYNK Wireless Transmitter SSH SYNKTX FCC Identifier VisionPro SYNK 26 Wireless LED Display SSH SYNKRX IC Regulations IC 4919C SYNKTX SYNK Wireless Transmitter IC 4919C SYNKRX VisionPro SYNK 26 Wir...

Страница 24: ...ffect the normal function of the display Do not use cables or accessories other than those provided with the display and transmitter as this may result in increased electromagnetic emissions or decreased immunity to such emissions If the display is used adjacent to or stacked with other equipment observe and verify normal operation of the display and transmitter in the configuration in which it wi...

Страница 25: ...reless Transmitter are intended for use in the electromagnetic environment specified below The customer or the user of the VisionPro SYNK 26 Wireless LED Display and SYNK Wireless Transmitter should ensure that they are used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge ESD IEC61000 4 2 8kV contact 15kV air 8...

Страница 26: ... according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol a Field strengths from fixed transmitters such as base ...

Страница 27: ...80 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz 2 0 3 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation 217 Hz 2 0 3 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation 217 Hz 0 2 0 3 9 5500 5785 Note Por...

Страница 28: ...ng Caution See Instructions for Use DC power control switch Wireless Transmission Country of origin Tested to comply with FCC Class B and C standards Storage and Transport Temperature Range Storage and Transport Humidity Range Alternating current Fuse Rating IC Industry Canada Federal law restricts this device to sale by or on the order of a physician Transmitter Medical Equipment is in accordance...