Service Information
12
HYDRAULIC TROUBLESHOOTING
Be sure the pedal linkage has been adjusted properly before beginning service on the jacks (see page 8 or
page 9).
PROBLEM/SYMPTOM
SOLUTION
Jack will not raise to full height.
Add hydraulic fluid (see p.14). Check for leaks.
Jack will not hold in raised position.
Close the needle valve completely. If the jack
holds, replace valve #1 (see p. 15). If the jack
does not hold, replace valve #2 (see p. 16).
Jack will not pump up and the jack actuator rod
does not
move.
Close the needle valve. If the jack will now pump
up, replace valve #1. If the jack still will not pump
up after closing the needle valve, replace valve #3
(see p. 16).
Jack will not pump up but the jack actuator rod
does
move when the pump pedal is activated.
Replace valve #2 (see p. 16).
Jack will not pump up and the jack actuator rod
may or may not move.
Remove excess air (vacuum) in system (see p.
16).
Contact Stryker technical service at 1–800–327–0770 for further assistance.
Содержание Thermaltek 1030
Страница 8: ...Notes 5 2 ...
Страница 15: ...Notes 10 2 ...
Страница 29: ...24 715 1 250 Side Control Base Assembly with Brakes ...
Страница 39: ...Notes 32 2 ...
Страница 41: ...Side Control Base Assembly with Jacks 34 Assembly part number 715 1 260 ...
Страница 45: ...Side Control Base Assembly With Jacks 37 1 Assembly part number 1210 2 10 ...
Страница 48: ...Notes 37 4 ...
Страница 49: ...End Control Base Assembly with Jacks 38 FOOT END HEAD END Assembly part number 716 1 270 ...
Страница 56: ...43 Notes ...
Страница 74: ...57 4 Notes ...
Страница 91: ...74 Notes ...
Страница 116: ...DH 2 97 1030 90 2 REV A 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...