INSTALLATION
(continued)
OBSERVING ALL SAFETY PROCEDURES, install a xenon bulb of the desired
wattage into the lamphouse. Check the bulb manufacturer’s documentation to confirm the recom-
mended current range of the 10 kW bulb (165-210 amperes DC). A new bulb is normally first oper-
ated at “nominal” current, which is around 85% of the maximum level (180 A.). DO NOT, at any
time, exceed the maximum current level specified by the bulb manufacturer.
IGNITE THE XENON BULB and check the current as indicated on the control
panel ammeter. Allow (30) seconds for the current to stabilize and provide an accurate reading. If
the current is not within the above-noted range, extinguish the bulb. It will be necessary to increase or
decrease the DC output.
OUTPUT CURRENT ADJUSTMENT
WARNING
Turn off ALL primary AC power before making any
adjustments or performing service procedures. Allow
several minutes for the capacitors to drain stored en-
ergy. The power supply normally operates warm to
hot; allow the unit to cool to safe temperature.
2