TROUBLESHOOTING
(continued)
Contactor does not energize (no audible “click”)
1. Line safety switch open. Turn “ON.”
2. AC line circuit breaker off. Turn “ON.”
3. Circuit breaker or fuse in AC line open. Check AC source.
4. REMOTE/LOCAL switch set incorrectly. Romote contact required when in REMOTE mode.
5. Lamphead interlock switch open. See lamphead manual.
6. Faulty K1 contactor coil or loose connection at coil terminals. Repair or replace.
7. Defective S1 thermal switch. Repair or replace.
8. Defective relay K1 or K6; Repair or replace.
Contactor pulls but lamphouse igniter does not fire
1. Faulty contactor contacts. With coil energized, check for continuity across the contacts from the
“L” side to the “T” side; repair or replace if defective.
2. Insufficient DC output. See INSTALLATION section; increase tap setting as required.
3. Faulty timer PC board assembly. If igniter fires when STRIKE switch is pressed, repair or replace
timer board.
4. Faulty igniter. See lamphead manual.
5. Low open circuit voltage (less than 110 V.DC).
a) Check ceramic resistors R1 & R2; should be in circuit and measure 100 Ohms.
b) Check boost diodes D2 and D3. See preceding DIODE TESTING section.
c) Check filter capacitors. Replace if defective.
Bulb requires multiple ignition pulses to light
1. Insufficient DC output. See INSTALLATION section; set tap setting as required.
2. Faulty or expired xenon bulb. Check for darkened envelope, worn electrodes; replace if required.
3. One or more faulty bridge diodes. See preceding DIODE TESTING section.
Bulb goes out during operation
1. Fuse F1 missing or open. Replace with same rated (3 ampere) fuse.
2. Excessive heat at rectifier heat sinks; thermal switch S1 opening. Check for free air flow, blower
operating at full speed. Check for loose connection.
3. Faulty lamphead interlock switch (“Door,” “Air,” etc.). See lamphead manual.
Excessive flicker in light output
1. Improper tap setting. All taps must be on same numbered or lettered step.
2. Faulty bridge diode. See preceding DIODE TESTING section.
3. Faulty xenon bulb. Check for cracked or sagging electrode(s).
4. Leaking xenon bulb. Check for white or bluish discolored envelope, high current and low voltage
(below 48 volts DC). Replace as required.