3-9
Installation Instructions
Uncrating/Installation Instructions
P-920010-163
b.Brown wire (L1) to L1 on TB1.
c.Blue wire (N) to L2/N on TB1.
d.Black wire (L3) to L3 on TB1.
NOTE: The black wire is not used. Be sure that it is properly insulated
in the junction box.
•
For 400 V, three phase, 50 Hz units:
a.Green/yellow wire (ground) to ground terminal.
b.Brown wire (L1) to L1 on TB1.
c.Black wire (L2) to L2 on TB1.
d.Black wire (L3) to L3 on TB1.
e.Blue wire to terminal identified X on TB1.
NOTE: The blue wire (neutral) is not used. Be sure that it is properly
insulated in the junction box.
3. Connect main supply cord to building electrical supply box.
Secure main supply cord with a strain relief (not by STERIS).
4. Re-install access panel on contactor box.