14.
Disassemble
communication
cable
connector
housing
to
expose
terminals
and
cable
connections.
Unplug
internal
connectors
from
each
other.
15.
Take
communication
cable
coming
from
washer
and
remove
insulation
and
strip
four
wire
leads
back
1”.
16.
Connect
four
wire
leads
to
spare
cable
connector/housing
from
service
kit.
Match
wire
colors
accordingly
(red/red,
black&shield/black&shield,
green/green,
white/white)
and
connect
to
mating
dosing
system
connector.