Sterilight Silver S12Q-PA Скачать руководство пользователя страница 15

To obtain this reading, press the push-button SWITCH a single time and follow the
steps as outlined on page 10, regarding the operation of this feature.

To obtain this reading, press the push-button SWITCH two times in succession and
follow steps as outline on page 10, regarding the operation of this feature.

2. 

Lamp life remaining (days):

3. 

Total days of operation:

Please refer to page 10 for explanation of this feature.
Note:  On the Silver “PLUS” systems, the audible tone provided for lamp failure is a
continuous alarm, rather than the intermittent (1 second on, 1 seconds off) 
condition on the basic Silver systems.

4. 

Lamp failure (blank screen):

5. 

Solenoid Output:

Working in conjunction with the UV intensity 
monitor, the Silver “PLUS” controller provides a 
powered, male IEC, solenoid (line voltage) 
connection (note: this is NOT a dry contact).  In 
addition, this solenoid connection is protected 
with a replaceable 2 amp isolated fuse.  When the 
UV intensity monitor senses that the water is not 
adequately being treated and drops to 49% UV 
intensity or below , the internal relay is opened 
thereby stopping AC power flowing to the 
normally closed solenoid valve.  The valve will 
remain closed (no power) until the UV level rises 
above 49%, at which time the solenoid will open, 
allowing for water to pass through.  To temporarily   
defer the operation of this solenoid output for up 
to 12 hours, please refer to the instructions outlined on page 11 of this manual.  

NOTE:

DURING BYPASS, THE 

“DO NOT CONSUME THE WATER”

tag included 

with this manual should be placed in a prominent location and the water should 
NOT be consumed until the system has returned to a safe condition.

Possible causes for low uv alarm conditions:

i) The UV lamp has perhaps reached a level whereby it can no longer adequately 

provide a sufficient level of disinfection due to age (> 9000 hours). The lamp 
should be replaced with a new lamp from the manufacturer of the same size and 
type.

ii) The quartz sleeve and/or the sensor window have become stained or dirty.  

Mineral deposits or sediment in the water that was not detected during the 
original water analysis may be the cause for this.

iii) Intermittent voltage drop in the household power supply reducing the lamp 

output. The lamp will return to normal when the power is restored to full 
voltage. Note: the monitoring system will not operate during power failures.

iv) The quality of the influent water has changed and is no longer within the 

acceptable operational range of the UV system. Perform a water analysis to 
determine the exact constituents and concentration levels.

SOLENOID INSTALLATION

12

Содержание Silver S12Q-PA

Страница 1: ...24 SSM 37 SSM 39 Installation Operation and Maintenance Manufactured in Canada by P N 520104 EPA 57987 CN 001 September 2005 425 Clair Road West P O Box 1719 Guelph ON N1L 1R1 Canada t 519 763 1032 f 519 763 5069 e water r can com i www r can com Owner s Manual ...

Страница 2: ...ce you have taken the first step in ensuring the safety of your water supply by using a totally non intrusive physical disinfection method Your Sterilight system uses the most advanced UV technology on the market and is designed to provide you with years of trouble free operation with minimal maintenance required ...

Страница 3: ...hematic Breakdown 1 Safety Instructions 2 Water Chemistry 3 Installing Your UV Disinfection System 3 5 Disinfection Procedure 6 7 Operating Maintenance Instructions 7 9 Operation 10 13 Troubleshooting 14 15 Silver Series Dose Flow Chart 15 Specifications 16 17 Manufacturer s Warranty 18 SYMBOLS ...

Страница 4: ...Controller POWER 100 240V 50 60Hz BA ICE S SILVER BASIC SYSTEM BA ICE SM SILVER PLUS SYSTEM IEC replacement power cords for Silver ICE Controller sold separately 260010 NORTH AMERICAN NEMA 5 15P 3 PRONG GROUNDED 260011 CONTINENTAL EUROPEAN CEE 7 7 2 PIN WITH GROUND SCHUKO 260012 UK VERSION BS 1363 3 PRONG GROUNDED 5 AMP FUSE 260013 AUSTRALIAN VERSION AS 3112 3 PRONG GROUNDED 260019 NO CONNECTOR 3 ...

Страница 5: ...ts not recommended or sold by the manufacturer distributor may cause an unsafe condition 8 Intended for indoor use only Do not install this disinfection system where it will be exposed to the weather or to temperatures below freezing Do not store this disinfection system where it will be exposed to the weather Do not store this disinfection system where it will be exposed to temperatures below fre...

Страница 6: ...ted to the stainless steel reactor chamber The disinfection system is designed to be mounted either horizontally or vertically at the point of use or point of entry depending on the specific flow rate of the unit Note The ideal installation is vertical with the lamp connector on top This is to prevent water damage from occurring on the lamp pins and lamp connector The ballast should be mounted eit...

Страница 7: ...outlined on page 6 of this document If the water is to be consumed while the system is off line the water must be boiled for twenty minutes prior to consumption 2 Select a suitable location for the disinfection system and its related components As it is recommended to install a ground fault protected circuit GFCI make sure that this is taken into consideration prior to any installation The system ...

Страница 8: ...ain to prevent any water from potentially entering the controller Affix the green ground wire to the grounding lug at the top of the reactor vessel and securely fasten with the lugnut provided Figure 1E 5 Install the UV lamp and UV sensor as outlined on pages 7 9 6 Install the over molded lamp connector as outlined on page 7 7 When all plumbing connections are made slowly turn on the water supply ...

Страница 9: ...d the chlorine disinfection solution you will need to leave the solution sit for a period of 20 30 minutes Reinstall the pre filter cartridge into the filter and then flush the chlorine solution from the system Make sure that each fixture that was disinfected in step two is completely flushed of the chlorine solution as the consumption of this water is not advised due to the extremely high concent...

Страница 10: ...n closing a seasonal home or leaving the unit in an area subject to freezing temperatures OPERATING MAINTENANCE INSTRUCTIONS Caution PRIOR TO PERFORMING ANY WORK ON THE DISINFECTION SYSTEM ALWAYS DISCONNECT THE POWER SUPPLY FIRST UV Lamp Replacement Cleaning 1 To replace the lamp there is NO need to disconnect the system from the water supply nor to drain the water from the reactor chamber Lamp re...

Страница 11: ... step 1 4 then perform the following steps a Shut off water supply and drain all lines b Remove the lowest connection on the disinfection system and drain the UV chamber use a small bucket under the unit to prevent a spill Note On S12Q PA SSM 39 systems the reactor is provided with a 1 4 drain port On this system simply remove drain plug and allow water to drain into a bucket c Remove aluminum gla...

Страница 12: ...e 4C until the sensor is free of the threaded sensor port 3 Once the sensor is free from the reactor chamber clean the quartz window with a commercial scale remover CLR or Lime A Way and a lint free cotton swab Figure 4D Follow all manufacturer s instructions regarding the cleaning fluid used Do not use an abrasive cleaner on the sensor window Scratching of the sensor window will void any manufact...

Страница 13: ...p connector 4 replace lamp connector 5 hold down the RESET switch while reapplying power to the controller 6 5 second delay will occur until you hear an audible tone LED display will read once again Once you hear the tone let go of the switch and the counter will be reset The delay switch is designed to allow you time to address the alarm while you obtain a new UV lamp Even though the alarm on the...

Страница 14: ... the UV level is this low to Indicates the UV level is nearing the point of unsafe UV intensity at this level the system should be CLOSELY examined to determine why the UV level is this low to Indicates the UV level has now reached a level that is unsafe At this level the water should not be consumed The system water supply should be examined to determine the reason for the low UV level of the UV ...

Страница 15: ...V level rises above 49 at which time the solenoid will open allowing for water to pass through To temporarily defer the operation of this solenoid output for up to 12 hours please refer to the instructions outlined on page 11 of this manual NOTE DURING BYPASS THE DO NOT CONSUME THE WATER tag included with this manual should be placed in a prominent location and the water should NOT be consumed unt...

Страница 16: ...ble comes with 20 meters 65 of cable for the 4 20 mA signal To install first remove the existing sensor cable from the Silver Plus controller Figure 5B and affix the new Y cable Figure 5C Next attach the male end of the existing sensor cable to the female end of the new Y cable Appropriately attach the 4 20 mA cable to the applicable equipment and ensure all connections are hand tight FIGURE 5A FI...

Страница 17: ...d by UV system Heated Product Water common problem caused by infrequent use of water run water until it return to ambient temperature Water Appears Milky caused by air in the water lines run water until air is purged Unit Leaking Water problem with o ring seal on gland nut and or UV sensor ensure o ring is in place check for cuts or abrasions clean o ring moisten with water and re install replace ...

Страница 18: ...o power lamp check to see if there is sufficient power to the UV system Low UV level displayed on screen test water supply to see if water quality meets recommended parameter limits clean quartz sleeve and sensor eye LED flashing A2 and then back to UV level low UV alarm deferral has been activated UV level has dropped below 50 and the audible alarm has been muted by pressing the reset switch and ...

Страница 19: ...MNPT Shipping Weight 2 3 kg 5 lbs 2 3 kg 5 lbs 2 7 kg 6 lbs 3 2 kg 7 lbs 4 5 kg 10 lbs 5 4 kg 12 lbs Electrical Voltage2 100 240V 50 60Hz2 100 240V 50 60Hz2 100 240V 50 60Hz2 100 240V 50 60Hz2 100 240V 50 60Hz 100 240V 50 60Hz Power Consumption 15 W 19 W 22 W 30 W 46 W 48 W Lamp Watts 10 W 14 W 17 W 25 W 37 W 39 W Maximum Operating Pressure 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi 8 62 b...

Страница 20: ... 2 kg 7 lbs 3 6 kg 8 lbs 5 0 kg 11 lbs 5 9 kg 13 lbs Electrical Voltage 100 240V 50 60Hz 100 240V 50 60Hz 100 240V 50 60Hz 100 240V 50 60Hz 100 240V 50 60Hz Power Consumption 19 W 22 W 30 W 46 W 48 W Lamp Watts 14 W 17 W 25 W 37 W 39 W Maximum Operating Pressure 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi Ambient Water Temperature 2 40 C 36 104 F 2 40 C 36 ...

Страница 21: ...nce 75 call factory for recommendations on applications where UVT 75 Where total hardness is less than 7 gpg the UV unit should operate efficiently provided the quartz sleeve is cleaned periodically If total hardness is over 7 gpg the water should be softened Warranty will be void if the proper steps are not taken to ensure that these impurities are not present 2 This limited Warranty shall not ap...

Страница 22: ... Public B Community Well C Your Own Well D Lake River E Cistern F Other specify ______________ 7 Do you have any other water treatment products specify __________________________________________________________________________________________ 8 Who installed your unit A Yourself B Plumber C Other 9 Unit purchase price __________ Installation cost __________ 10 Do you have or use bottled water A Ye...

Страница 23: ...amp Here DO NOT CONSUME THE WATER The system has been placed on by pass and as a result the water in this distribution system should not be consumed until the system has been fully sanitized and returned to service Please refer to the Installation Manual for complete disinfection procedure ...

Отзывы: