THIS SECTION IS FOR ENGINEERING USE ONLY
❑
Draining Fault
The unit is not draining correctly.
Check that the drain hose is installed correctly.
Check that the drain pump is running.
Check that the drain hose is not blocked.
Check the zone valve is working.
If the drain is restricted by the hospital’s drainage system, the max drain time might need increasing.
❑
S180 Not Heating up
Perform heater reset as per Maintenance section of this manual.
❑
No Fill Fault
The bath has failed to reach the correct level in time. Check that the water supply is turned on. Check
levels are set correct. Check inlet filters on the solenoid valves are not blocked.
❑
Power Loss Fault
The power supply to the pasteuriser has been lost during a cycle.
Issue 5 4/12/2019
34
Engineer Fault Finding
Содержание S180
Страница 1: ...Operating Manual for S180 Pasteuriser Serial No This Manuals belongs to ...
Страница 4: ...Please Add Specification Issue 5 4 12 2019 4 ...
Страница 11: ...250ml 130ml 100ml 50ml Issue 5 4 12 2019 11 BOTTLE CHOICES Induction Sealer ...
Страница 12: ...BOTTLE CHOICES 250ml 130ml 100ml 50ml Issue 5 4 12 2019 12 Pasteuriser Bottles ...
Страница 15: ...HOME SCREEN Issue 5 4 12 2019 15 Pasteuriser Operation ...
Страница 17: ...Bottle Probe Position Issue 5 4 12 2019 17 Pasteuriser Operation ...
Страница 29: ...Data Logger Operation LIST VIEW GRAPH VIEW Issue 5 4 12 2019 29 Data Logger Operation ...
Страница 40: ... Please add breakdown form Breakdown report form Issue 5 4 12 2019 40 ...
Страница 41: ... Please add breakdown form Breakdown report form Issue 5 4 12 2019 41 ...
Страница 42: ...Please add sequences of events chart Contact Us ...
Страница 43: ...Contact Us Please add Graph ...