STEINDORFF

®

 

 

8.

 

Illumination: Transmitted 6V-20W light source, brightness adjustable. 

9.

 

Line Power: 220V/50-60Hz. 

10.

 

Binocular Viewing Head 

11.

 

Compensator: Quarter Wedge 1-4 Order, Full Wave 1l Gypsum & Quarter Wave 1/4l Mica. 

12.

 

Condenser: Abbe two-lens condenser, N.A. = 1.25. 

 

Unpacking 

Identify the version of the Microscope purchased, unpack and bring the Microscope out of the case. Check up the 

accessories in the case. 

 

Operation 

Insert the plug of the power of this microscope into a suitable grounded 220V AC line power socket. Turn on the 

power by the switch & the indication light will be lit. Switch ON the Microscope Illuminator. 

1)

 

Place a specimen onto the stage. Make an observation, pay attention to pull the lever for pulling the 

analyzer out from the light path. Open the Iris Diaphragm. Make sure that the field Iris Diaphragm 

coincides with the field of view. 

 

2)

 

Secure the specimen properly with the clips on the Rotary Stage. 

 

3)

 

Adjust the center of the stage using the stage centering screw. 

Focus the specimen sharply. Find a marked feature in the field 

of view and make it to situate at the center of the ocular’s cross 

hairs. Rotate the stage, if the optical axis of the objectives not 

coincides with the center of the rotation of the stage, then the 

target  selected  will  rotate  about  certain  center  S  (that  is  the 

rotating center of the stage), the trajectory of which is a circle. 

Turn the target point to the point 01 to move toward the point 

S  and  coincide  with  it.  Then,  turn  the  stage  again,  to  view 

whether  the  two  points  coincide  with  each  other,  if  any 

deviation exists, repeat this procedure again. 

 

 

0

1

 

Содержание NYMC62B00

Страница 1: ...STEINDORFF NYMC62B00 Polarizing Microscope ...

Страница 2: ...tage Height Adjustment Screw 7 Coarse Focusing Knobs 8 Fine Focusing Knobs 9 Condenser Centering Screws 10 Condenser Focusing Knob 11 Brightness Control 12 Power Switch 13 Power Indicator 14 Light Collector 15 Graduated Marks of Polarizer 16 Iris Diaphragm Lever 17 Stage Centering Screws Detachable 18 Nosepiece 19 Graduated Marks on Analyzer 20 Analyzer Lever ...

Страница 3: ...ications 1 Observation Tube Binocular Head 2 Objectives stress less achromatic objectives Magnifying Power Objectives Numerical Aperture N A Thickness of Cover Glass 4x 0 10 10x 0 25 20x 0 40 40x 0 65 0 17 60x optional 0 85 0 17 3 Eyepieces Type Magnifying Power Standard Eyepiece Pair 10x Eyepiece with Crosshairs 5x Eyepiece with Graticule 10x 4 Final magnifying powers of objectives and eyepieces ...

Страница 4: ...he lever for pulling the analyzer out from the light path Open the Iris Diaphragm Make sure that the field Iris Diaphragm coincides with the field of view 2 Secure the specimen properly with the clips on the Rotary Stage 3 Adjust the center of the stage using the stage centering screw Focus the specimen sharply Find a marked feature in the field of view and make it to situate at the center of the ...

Страница 5: ...d for high power examination Insert Bertrand lens into the light path when using orthogonal polarizing light The swinging upper lenses of the condenser are arranged so that they may be moved into the optical axis of the microscope for examining the conical characteristic of the specimen 7 When an indistinct image of the filament of the bulb reveal in the field of view add a ground glass properly W...

Страница 6: ...ismantle any part of the instrument If there is any dirt on the surfaces of the lens wipe it off gently with some absorbent cotton moistened with xylene Dust on the lens surfaces can be blown off by using an air blower 2 Oil the mechanical moving parts of the instrument with some watch or clock lubricant 3 Avoid contacting high temperature 4 Objectives and eyepieces should replace back to their co...

Отзывы: