EN
Espire Elleboog Pro & Hybrid van Steeper Group, distributeur Loth Fabenim B.V.
Espire
TM
Elleboog Pro & Hybrid Instruction Manual - For Prosthetists
22
Inputs
3
Note: Due to a taller input connector, the Coapt Gen2 system requires a specifically-
shaped strain relief disc. The Coapt EMG Interface Cable Connector increases the
elbow build height by 0.15”/3.8mm.
Coapt Strain Relief Disc
1. Thread the COMPLETE CALIBRATION Button
wire ribbon through the slotted hole of the
disc, then attach the connector to the board.
2. Insert the EMG Interface Cable connector
through the matching shape on the disc, then
attach to the board.
3. Fasten the wires to the strain relief disc using
the two provided zip-ties. This is necessary
4. to prevent accidental wire disconnection,
twisting, and damage.
5.
Use (two) M2 x 5mm fasteners and washers
to secure the disc to the input board. A 1.3
mm Allen wrench has been provided to
tighten the fasteners finger-snug (2 in-lb/0.23
N-m). It is not necessary to seal the input
board.
COMPLETE
CALIBRATION
Button Wire
EMG Interface
Cable Connector