STATUS SCIENTIFIC CONTROLS
FGD3 Infrared Gas Detector Heads
TD06/024
Issue:
7
Change Note:
1904
Page 23
5.2.12
C: 19 - Positive Zero Suppression
This option is used to allow the user to suppress small amounts of positive sensor zero
drift. The setting can be set between 0 and 10% of the sensor range as set by the FSD
value.
•
Press MENU to open the menu system.
•
Using the NEXT and PREVIOUS buttons, select menu option:
C:19
•
Press ENTER.
•
Using the INCREASE and DECREASE buttons, set the required zero
suppression value.
•
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits
without any change.
•
Press MENU to close the menu system.
5.2.13
C: 20 - Negative Zero Suppression
This option is used to allow the user to suppress small amounts of negative sensor
zero drift. The setting can be set between 0 and 10% of the sensor range as set by the
FSD value.
•
Press MENU to open the menu system.
•
Using the NEXT and PREVIOUS buttons, select menu option:
C:20
•
Press ENTER.
•
Using the INCREASE and DECREASE buttons, set the required zero
suppression value.
•
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits
without any change.
•
Press MENU to close the menu system.