7.5 ParametersOverview
Parameters
Factory setting
Value range/Selection
Probe ID
Numeric
various
1
... 99999
Transmission
interval
15
s
1 ... 3600 s
Transmission
frequency
868.4 MHz
868.4
MHz
Display only, cannot be edited!
Probe type
RTD
temperature
probe
Potentiometer,
voltage,
thermocouple,
RTD temperature probe
Sensor
Pt100
depending on the probe type,
Pt100 DIN
EN 60751 -200 ... +600 °C
Connection
type
3-wire circuit
3-wire circuit
, 2-wire circuit
Lead
resistance
0 Ohm
(
0
to 22) Ohm Only with 2-wire circuit!
Resistance RP
10000 Ohm
(5 to
10000
)
Ohm Only for potentiometer
probe type!
Resistance R0
0 Ohm
(
0
to 4000) Ohm Only for potentiometer
probe type!
TAG number
10-digit number, freely selectable
Information
text
10-digit number, freely selectable
Installation
date
current date
any date
TIP!
At a transmission interval of > 15 s, the probe transmits a so-
called link telegram (after a set-up transmission), i. e. the
telegrams are transmitted at the factory-set interval of 15 s
for a period of 30 minutes, and only then at the set
transmission interval.