background image

Ø 3mm

Screw Driver

20

20

20

20....000    

000    

000    

000    

Digital mA meter

Power supply (10 to 30) Vdc

Resistor Decade box

User adjust function allows manual adjustment of the output current, 
this  is  useful  for  minor  calibration  adjustment  or  trimming  out  any 
sensor error, ± 5% of range adjustment is available at both offset and 
span.  Raise and lower buttons are provided on the front panel of the 
transmitter,  accessed  using  a 3 mm flat blade screw  driver. Insert  the 
screw  driver  into  the  appropriate  slot  to  operate  the  button.  The 
button has a click action. 

METHOD
1.0    In  USBSPeedLink  software  set  the  “Buttons”  drop  box  to 
trim (this is the default setting).
Connect    transmitter  to  a  suitable  Resistor  decade  box  or 
sensor. Connect output to a dc supply, connecting a digital mA 
current  meter  in  series    with  the  output.    Turn  supply  on,  set 
input to either offset or span calibration point. 

+

-

20

20

20

20....000    

000    

000    

000    

+

-

SEM1605P

or

23.0 

Adjust 

output 

current 

by 

pressing 

either the raise or lower 
button,  single  click  to 
step  advance  ,  or  press 
continuously  to  auto 
advance.

Raise

Lower

34.0 

Once 

adjust 

is 

complete 

allow 

30 

seconds  with  no  button 
press, the transmitter will 
time  out  and  return  to 
normal operation.

D2588-01-03  CN5602 SEM1605P User Guide

> 30 s

The 

transmitter 

will 

automatically  detect  the 
correct  adjust  point  (offset 
or  span)  based  on  the 
output current drive. Offset 
will  be  adjusted  when  the 
current    is  between  (3.8  to 
6)  mA,    span  when  the 
current  is  between  (18  to 
22)  mA.  No  trim  action 
occurs at any other current.

offset

span

3.8

6

18

22 mA

 or sensor

ADJUST

STATE

ADJUST

STATE

ADJUST

STATE

Button Configuration options.

The SEM1605P has three options selectabe in the USB Speed Link software, Trim, Range and off

Ω

Ω

Ø 3mm

Screw Driver

Power supply (10 to 30) Vdc

Resistor Decade box

User range function allows manual adjustment of the 4 mA and 20 mA 
output range in relation to the input value.

Method  In USBSPeedLink software set the “Buttons” drop box 
to active range (this is not the default setting).
1.0 Connect the resistor decade box or the input sensor to the 
SEM1605P  using the  three connection terminals.   Connect the 
SEM1605P to a (10 to 30) VDC power supply, a digital ammeter 
connected  in  series  with  the  SEM1605P  will  be  useful  to 
monitor the (4 to 20) mA current but is not essential.
Turn on the supply and allow 1 minute warm up period.

2.0 Set the resistance decade box to the equivalent resistance 
of the sensor for the required low range temperature, or apply 
required low range temperature to the sensor.
Allow 10 seconds for the SEM1605P to settle.

3.0 Press and hold the Low range ADJUST button, marked with 
a  blue  down  arrow,  until  the  STATE  LED  starts  to  flash,  then 
release the button.
Press  and  release  the  Low  range  ADJUST  button  again,  the 
STATE  LED will flash  quickly  for a  short  time  and  the  new low 
range will be stored. The output current will go to 4.0 mA.

4.0 Set the resistance decade box to the equivalent resistance 
of the sensor for the required high range temperature, or apply 
required high range temperature to the sensor.
Allow 10 seconds for the SEM1605P to settle.

5.0 Press and hold the High range ADJUST button, marked with 
a  Red  up  arrow,  until  the  STATE  LED  starts  to  flash,  then 
release the button.
Press  and  release  the  High  range  ADJUST  button  again,  the 
STATE LED will flash quickly for a short time and the new high 
range will be stored. The output current will go to 20.0 mA.
The ranging of the SEM1605P is now complete.

Note:- The Low and High user adjust can be set individually and 
in any order as required.

+

-

+

-

SEM1605P

or

 or sensor

Ω

Ω

Pressing the button has no action.

USER ADJUST

USER RANGE

OFF

2.0  Press  and  hold  the 
red  up  arrow  for  10  s 
until  the  STATE  LED 
starts  to  flash,  then 
release the button