background image

D2614-01-03 CN5819 

5.1~USB CONFIGURATION 

Using the USBSpeedLink software to program a new configuration into the device 

work from left to right through the option tabs. 

The USB port is located inside the housing on the rear panel. 

Configuration steps 

Install and run the USBSpeedLink software 

Select the correct programming page in the software (follow any on screen 

safety precautions) 

Connect the USB lead to the PC, connect the USB lead to the device 

Read the unit configuration into the software 

Re-configure or adjust configuration options as required 

Send the new configuration to the unit 

 

5.1.1~Sensor option tab 

Use this tab to pre-set RH and temperature values; this can be used for system testing 

and diagnostics. 

 

5.1.2~Relay tab (showing live readings in the “Data” section) 

Use this tab to set the relay controls. 

 

 

Relay options for %RH, temperature, dew point and delta t. source 

Set point 

Dead band   

Low alarm 

Low control (inverted low alarm) 

High alarm 

High control (inverted high alarm) 

Alarm inside band 

Alarm outside band 

Low battery alarm 

Alarm off 

A diagram will display to show alarm type function 

 

5.1.3~Basic display options tab 

Use this tab to select the display value options, °C or °F and LED function 

 

 

!

 

 

 

 

 

 

 

5.1.4~Advanced display options tab 

Use this tab to set up the advance messaging options. 

 

5.1.4~Advanced display options tab (continued) 

6 user set ranges/bands can be set for either %RH, temperature, dew point or delta t. 

Each range can have 1 or 2 messages assigned that will display on the screen when 

the input source signal is within the set band. Also, the LED can be set to flash, within 

the selected range. 

Up to 7 user messages can be entered or pre-set messages can be used. 

Relay in or out of alarm trip can also be used to prompt specific messages to be 

displayed. 

 

5.1.5~Other display options tab 

Use this tab to set the “Tag” and “Contact” free type fields stored to the device. A 

passcode to protect the device log from being overwritten by an Android device can be 

added.  

Longitude and latitude data can be entered, this will show in the log files and can help 

identifying data logs recorded from multiple devices.  

 

6~DATA LOGGING 

USB and Android NFC software allows the user to read existing logs, change the log 

manifest, start a new log, synchronise the instrument clock and reset the 

maximum/minimum readings.  

Logs can run to a fixed number and stop or continually roll over, up to 5000 log points 

can be recorded. The start of the log can be delayed up to one month.  

Data from can be viewed on screen as a graph or table and can also be saved to a 

csv file. 

 

6~1 USB LOGGING 

 

 

 

 

 

USB LogLink software 

 

 

6~2 NFC LOGGING 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Status Instruments Ltd, Status Business Park, Gannaway Lane, Tewkesbury, 

Gloucestershire, UK, GL20 8FD, 

 

Web Page: www.status.co.uk, Email: [email protected] 
Technical Support: [email protected] 
Tel: +44 (0) 1684 296818, Fax:  +44 (0) 1684 293746 

USB LOGGER INTERFACE (connector inside housing) 

The USB interface allows the instrument to communicate with a PC running 

the USBLogLink software  

The prime function of the interface is to read logged data from the device 

using free software available to download.  

The software allows the user to read existing logs, change the log manifest, 

start a new log, synchronise the instrument clock and reset the 

maximum/minimum readings. 

USBLogLink is available from the manufacturer or supplier.  

 

 

NFC LOGGER INTERFACE   

The NFC interface allows the instrument to communicate with an Android 

device using NFC connectivity.  

The prime function of the interface is to read logged data from the device 

using a free app, which is available for downloading to Android devices.  

The app allows the user to read existing logs, change the log manifest, 

start a new log, synchronise the instrument clock and reset the 

maximum/minimum/average readings. Logs can run to a fixed number and 

stop or continually roll over, up to 5000 log points can be recorded. The 

start of the log can be delayed up to one month. 

Note:

 For larger logs the data may take over a minute to fully download via 

the NFC interface. 

Note:

 Not all Android devices fully implement the near field communication 

(NFC) standard and may not hold the full 5000 log points. 

Download Status Instruments apps from the Google Play Store 

 

 

MULTIFUNCTION ALERT LED  

The alert LED normal state is off, on alert the LED will emit an 

intense white light pulse every 5 seconds. The LED can be 

programmed to pulse on any of the following combined events: 

Mode 

Description 

No events   The LED never operates, extending battery life. 

(Factory default setting) 

Battery 

Alert on low battery detect. 

Trip  

Alert when relay trip is on. 

Temperature  In advanced mode only the alert LED can be made to 

alert in any one of eight user-set temperature bands. Example, to 

alert operator  when temperature is outside a safe operating range.  

The function of the alert LED can be further enhanced with the option 

of displaying an alert message in advanced display mode. 

 

 