D2614-01-03 CN5819 

 

DM650HM USER GUIDE 

BATTERY POWERED %RH, °C/°F METER 

WITH RELAY TRIP AND DATA LOGGING FUNCTION

 

Important - Please read this document before installing. 

Every effort has been taken to ensure the accuracy of this document; however, we do 

not accept responsibility for damage, injury, loss or expense resulting from errors and 

omissions, and we reserve the right of amendment without notice. 

IMPORTANT – CE, UKCA & SAFETY REQUIREMENTS

 

 

The instrument is designed to be directly attached to the sensor probe assembly. 
Remote Probes may be used but the user must ensure all sensor and cable entries 
maintain environmental protection to at least IP65 rating.   
To maintain CE EMC requirements, input wires must be less than 3 metres. 
The product contains no serviceable parts, or internal adjustments. No attempt must 
be made to repair this product. Faulty units must be returned to supplier for repair. 
This product must be installed by a qualified person. All electrical wiring must be 
carried out in accordance with the appropriate regulations for the place of installation. 
Before attempting any electrical connection work, please ensure the battery is 
removed 
 

ABSOLUTE MAXIMUM CONDITIONS (To exceed may cause damage to the unit). 
Battery Voltage 

± 3.7 VDC (Protected for over-voltage and reverse connection) 

 

 

Ambient 

Temperature (-30 to 70) °C, Humidity (10 to 95) % RH (Non-
condensing) 

Relays 

50 VDC 40 VAC RMS 

 
 

Important – Read this 

document before installing.

 

--------------------------------------------------------------------------------- 

1~DESCRIPTION

The DM650HM is a battery powered indicator with built-in sensors designed to 

display % relative humidity, temperature, dew point and delta temperature. The 

instrument offers an advanced display-mode, allowing the user to display not only 

process variables but one or two (1–32) character messages. Additional to the 

messages, the user may select an alert event to occur when the process is within a 

band. The change-over relay can be used together with the messaging and alert to 

provide switching at user-set points. In addition to this, on board datalogging is also 

available to record process details at up to 5000 points. 

 

2~RECEIVING AND UNPACKING

Please inspect the packaging and instrument thoroughly for any signs of transit 

damage. If the instrument has been damaged, please notify your supplier immediately. 

 

3~SPECIFICATION. 

Refer to data sheet for full specification. 

Factory default 

Displays %RH, °C, relay not set 

 

4~INSTALLATION AND WIRING

  

4.1~MECHANICAL.  

The instrument is a high accuracy digital thermometer. In order to ensure correct 

operation, the following must be observed: 

The product must be stored in a dry clean environment and remain in original 

packaging prior to installation. 

The instrument must not be installed adjacent to electro-mechanical starters, 

controllers, thyristor power units or electrical switch gear. 

Any cleaning of the instrument must be done using a mild detergent and soft cloth. No 

solvents or abrasive cleaners should be used. 

Stated ambient operating conditions must not be exceeded. Battery life will reduce 

with higher ambient temperature operating conditions. 

 

4.2~ELECTRICAL. 

For a wiring diagram please refer to the rear panel of the DM650HM inside the 

case housing.  

Connections: NO (Normally Open), NC (Normally Closed), and C (Common). 

Two-part connectors are used for input and relay connections, allowing the unit to be 

easily removed from the housing for reprograming or data download if this is not 

possible in situ. On insertion of battery the unit will run through a power up check; 

during this time the relay may change state. 

IMPORTANT: Always remove battery before any wiring takes place. Gain access to 

the connectors and battery holder by unscrewing the cap ring to release front panel 

assembly from case.  

RELAY CONNECTION: A relay with changeover contacts is available. Screw 

terminals are provided for connection for wire size 16 to 20 AWG. The relay contacts 

are rated at 48 VDC 28 VAC RMS @ 1 A (5 mA minimum current) see DM650HM 

data sheet. 

BATTERY: To remove battery, use screwdriver to ease the positive end of the battery 

out of holder. Insert new battery negative end first then press into place. (Observe 

polarity). Battery type 3.6 V Lithium (2.4 A/Hr) CR14505 (IEC) AA case style. Please 

dispose of the battery in a responsible way. 

 

5~USER CONFIGURATION. 

 

 IMPORTANT

 

READ COMPLETE SECTION BEFORE ATTEMPTING CONFIGURATION.  

 

The instrument is provided with a USB interface for direct connection to a PC. Free 
software USBSpeedLink is available. Please refer to the USBSpeedLink software for 
further information on configuration. The software can be downloaded from 
www.status.co.uk.  
DISPLAY: The display provides six 14 segment characters for display of temperature 
and alpha-numeric messages, together with an 8 segment bar graph and six icons. 
The display can operate in an ambient temperature range of (-30 to 70) °C, but at 
temperatures lower than -5 °C (due to the slow LCD speed) scrolled messaging is not 
practical. 

9

1

2

3

4

6

8

7

8

 

1.  NFC - The Symbol is on when a NFC field is detected. When a detected field is lost the symbol 

will turn off after a few seconds. 

2. TRANSMIT/RECEIVE: Symbol on when either NFC or USB communication is active. 

3. USB: Symbol on when USB port is connected to a PC. Please note battery is not required during 

configuration. 

4. LOG: Not used 

5. Not used 

6. BAR GRAPH: Indicates the state of the logger. The condition is dependent on the selected 

logger mode, either Single or Rolling mode. 

 

Single Mode (Log to the maximum number of logs then stop) 

 

 

                 Start of Log                                    Mid Log                            End of Log (alternating) 

 

 

 

Rolling Mode (Log to the maximum number of logs then as each new log is taken the oldest log is 

discarded) 

                  Start of Log                                 Mid Log                    Rolled over Log (max bar toggles) 

 

 

7. DEG: Degree symbol “°” used to indicate either °C or °F on the last digit. 

8. DIGITS: Six-digit 14 segment display with – sign, range 9999.9 to -9999.9. Advanced mode offers 

two temperature-dependent 32-character message options. 

9. WARNING ICON: This symbol will toggle on and off to indicate a warning. The warning symbol will 

be active either when the sensor signal is out of range, not connected or when the battery is low 

 