Static Solutions Inc. - Ohm - Stat RT - 1000 Information
5
Resistivity Test Ranges
1 X 10
3
– 9 X 10
6
ohms @ 10 volts
1 X 10
6
– 9 X 10
12
ohms @ 100 volts
Power- Battery Supply
A 9 volt battery is supplied. If a
rechargeable battery is used, the
recharging can be done inside
the meter.
The 12 volt power supply is
both a adapter and a recharging
device. With a 9 volt alkaline
battery the expected life is
approximately 30 hours or 5000
twenty second measurements.
Using an adapter not designed
for this meter may cause serious
damage negating the warranty
on the meter. Only a 110 volt
DC, center positive 200 mA
adapter is recommended.
A low battery indicator will
alert the user to replace a weak
battery.
If a weak battery is not replaced
the meter may continue to give
test readings, but these readings
should not be considered
accurate.
Operating Conditions :
32
0
F-100
0
F (0
0
C-38
0
C)
Display:
Two inline, thirty two .20" characters are
alpha numeric. The LCD displays
all test results simultaneously.
Meter Weight :
10.2 oz. (289 mg.). Meter and battery
only.
12 lbs. (5.476 kg
. )
including meter,
adapter, probes, leads, and case.
Dimensions :
7.50 inch ( 19.05 cm ) L
4.00 inch (10.16 cm) W
1.5 inch (3.81 cm ) H.
Probes:
Two 2.87 inch (7.2 cm)
long parallel, conductive
rubber re-placeable probes
on the bottom of the meter
case. They conform to
ASTM, ANSI, and CECC
standards.
Two 5 pound (2.27 kg.),
2.5
inch (6.4 cm) diameter ,
conductive Shore A 50-60
durometer external rubber
external probes-enclosed.
Two smock clip test
probes- are enclosed
One 8 inch X 12 inch
stainless steel probe to be
used to test chairs and
shoes are enclosed
One ground tester to
check on the electrical
ground is enclosed
OPTIONAL
ACCESSORIES.
One 5 pound concentric
ring probe conforming to
EOS/ESD, ANSI and CECC
standards.
One 2 inch diameter , 6
inch long, dual probe to be
used to test the electrical
resistance of shoes and
wrist strap readings.
One miniature resistance
probe.
Current Limit:
1 milliamp DC current for 10 volts
0.1
milliamp DC current for 100 volts.
Accuracy:
10
3
– 10
8
ohms +/-10%
10
9
-
10
10
ohms +/-15%
10
11
- 10
12
ohms +/-25%
Zero:
Automatic zeroing.
Power Switch
:
Momentary on, press button
Automatic zeroing
Operating Conditions :
32
0
F-100
0
F (0
0
-38
0
C)