Check your display and printed results during operation. Your display and printer provide 
useful information such as the values you enter, the mode of operation you select, whether or 
not you remembered to blank, and all absorbance readings. Monitoring the printer and display 
during operation may help you detect an error in the making. 

Check the calibration and linearity of your instrument against some standard reference 
periodically. DRI-DYE

®

 Check Strips are recommended. 

Appropriate controls should be run with each assay as indicated in the package inserts of the 
chemistry products used. If controls do not give expected results, the assay is invalid. 

3.3 Maintenance

It is important to follow the installation instructions carefully, using only a suitable power supply 
and placing the instrument with the proper clearance for good air circulation around it. Excessive 
vibration should be avoided. During shipments, use the original packing material or other 
suitable protective foam. 

This reader is essentially a maintenance free instrument. To ensure maximum trouble free 
operation, the instrument need only be kept dry. Although the filters are well sealed, their life 
may be decreased if the instrument is maintained in a very humid environment (greater than 
80% humidity). Using the instrument in an air conditioned room is recommended for humid 
climates. Extreme temperature shock is also harmful to the filters. Maximum changes of 5ºC per 
minute are recommended especially at the lower limits of -50ºC, where permanent damage may 
occur. The upper limits of adequate performance for the filters are 120ºC. The instrument is 
designed for use at 20ºC and performs according to specifications in the range of 18-35ºC. 

Cleaning should be done only when obviously necessary. Use a dry cloth or duster to remove 
dry dust and dirt. Use only a slightly damp, soft cloth to clean up spills. Water or 70% 
isopropanol may be used to dampen the cloth. Use of other chemicals or abrasive scrubbing 
may damage the cover. Be careful not to soak liquids up under the keyboard overlay. If this 
happens, simply allow the keyboard to dry before resuming operation. Such a spill does not 
present a hazard, but may cause temporary malfunction of the keys. 

Содержание 303 Plus

Страница 1: ...Copyright 2001 Doc 303 Revised 04 2007 Rev J Stat Fax 303 Plus Operator s Manual...

Страница 2: ......

Страница 3: ...g a Strip Format 16 2 1 2 Selecting a Mode 17 2 1 3 Quitting a Mode 18 2 1 4 Selecting Filters 18 2 1 5 Selecting a Blank 19 2 1 6 Selecting Duplicates 19 2 1 7 Unit Codes 20 2 1 8 Selecting Cutoff In...

Страница 4: ...bration and Linearity 36 3 0 Additional Tips and Information 37 3 1 Restoring Electronic Calibration 37 3 2 Operating Precautions 37 3 3 Maintenance 38 3 4 Troubleshooting 39 3 5 References 41 4 0 App...

Страница 5: ...ates concentrations and prints results It comes with a built in thermal graphics printer with the capability to plot standard curves Besides providing absorbance readings this instrument also offers p...

Страница 6: ...e design with a timed lamp saving feature further assures the reliability of this instrument 1 2 Principles of Operation Light energy from an overhead lamp is focused by a lens directed through an ape...

Страница 7: ......

Страница 8: ...strument may be impaired Use Proper Power Cord Use only the power cord specified for this product and certified for the country of use Ground the Product This product is grounded through the grounding...

Страница 9: ......

Страница 10: ...the unit should be turned off Look at the rear panel of the instrument to check that the power switch is in the Off position Power Cord Requirements Use only the power cord specified for this product...

Страница 11: ...normally if the power supply is interrupted If this occurs turn the instrument off for a moment When you turn the instrument back on it will resume normal operation but a standard curve which was not...

Страница 12: ...cut a new straight edge before trying again Drop the roll of paper into the printer paper well in back spooling up any loose paper Replace the paper cover by pinching its sides together and placing it...

Страница 13: ...crostrip reader More details on the operation of each feature are provided in section 2 Operating Procedures The voltage select switch is on the base The power switch and serial connector are on the r...

Страница 14: ...ptake Mode Pressing the INTRP key causes the reader to ask for positive and negative cutoff points to be used for automatic sample interpretation Pressing the CNTRL key causes the reader to ask for th...

Страница 15: ...total number of wells to read and print The key assignments for the filters are as follows For all models key 1 is 405nm key 2 is 450nm key 3 is 492nm key 4 is 630nm 6 filter UV key 5 is 545nm key 6 i...

Страница 16: ...CD 16 characters 5x7 dot matrix Internal Printer Thermal dot matrix 20 characters per line Keyboard 16 key membrane switch 4X4 Calculation Modes Single point calibration Multi point calibration with r...

Страница 17: ...supplier Humidity 80 for temperature up to 31 C decreasing linearity to 50 humidity at 40 C Mains supply voltage fluctuations not to exceed 10 of nominal voltage Recommended Environmental Conditions...

Страница 18: ...14 Stat Fax 303 Plus Operator s Manual Rev J Awareness Technology Inc Notes...

Страница 19: ...o aid the user this instrument may be operated very simply in most modes by bypassing optional features For every test the instrument will require a mode selection and filter choices Thereafter only t...

Страница 20: ...lections If the lamp warm up is not completed by the time you press the ENTER key to read the first strip the display will indicate LAMP WARM UP 00 where 00 represents the number of seconds remaining...

Страница 21: ...gle point standard curve passing through the point 0 0 A blank is required in the first well or first two wells if duplicate blank is selected A factor equal to the concentration of the calibrator the...

Страница 22: ...entration 2 1 3 Quitting a Mode If a test is to be saved in the user test menu this must be done before canceling the mode See section 2 3 1 for instructions on Storing a User Test To quit a mode pres...

Страница 23: ...ank and calibrators When duplicates are chosen the mean absorbance reading is used as the reference absorbance for the blank and calibrator You may also choose to read specimens singly or in duplicate...

Страница 24: ...utoff When a single upper cutoff is used any concentration which is equal to or greater than the cutoff will be labeled as positive The letters P and N are shown in the right most column of the printo...

Страница 25: ...the 8 well mode the first set of wells right most carrier position will be Strip 1 row A through H If you are in the 12 well mode the first set of wells right most carrier position will be Row A stri...

Страница 26: ...eak apart wells in the 12 well carrier start from the front of row A and fill all 12 locations then begin at the front of row B etc and continue until done Ensure that the wells are firmly placed in t...

Страница 27: ...y for the next strip the display will indicate Set Carrier Then Press Enter To continue reading in the Absorbance mode position the carrier for the next strip and press the ENTER key Since there are n...

Страница 28: ...can have any positive or negative numerical value including 0 and 1 Press YES After the filters are selected if you chose no differential the Offset Absorbance questions will follow see section 2 4 5...

Страница 29: ...w mNeg C mNeg C Type in each value and press ENTER The display will show Num of PCs Type in the number of Positive Controls and press ENTER The display will show Pos C Pos C Type in each value and pre...

Страница 30: ...same order described in the regular cutoff mode NOTE Be sure to read the and signs in each display because blank positive control and negative control parameters are entered in a reverse way when usin...

Страница 31: ...e in the number of Cutoff Controls and press ENTER The display will show C O C C O C Type in each value and press ENTER The display will show Num of PCs Type in the number of Positive Controls and pre...

Страница 32: ...to access the regression mode Regression Y N will be displayed Answer YES This mode calculates a best fit linear equation based upon the standard points It is designed for use with either linear loga...

Страница 33: ...right most carrier position and press ENTER While the strip is being automatically positioned for reading the display will indicate Positioning Strp After the strip or strips containing the standards...

Страница 34: ...est 2 calibrator points The calibrators absorbances must be either decreasing or increasing down the strip i e if the first calibrator is the darkest the second must be the next darkest etc and if the...

Страница 35: ...e is obtained or the mode is canceled To read and calculate specimens answer NO to the question DELETE WELLS Y N The instrument will be ready for the next strip when it displays Set Carrier Then Press...

Страница 36: ...Enter To continue reading in the Uptake mode load the next strip and press the ENTER key Press CLEAR twice to cancel the mode 2 3 User Test Menu This instrument offers the operator 8K bytes of non vol...

Страница 37: ...u will see DELETE TEST Y N Answer YES When you see DELETE TEST in the display again type in the number of the test you wish to delete and press the ENTER key The instrument will display DELETE TEST Y...

Страница 38: ...e lost when the instrument is turned off or unplugged The time will be displayed whenever the instrument displays Ready The time and date will be printed on all assay reports on either printer The dat...

Страница 39: ...ORY IS FULL is displayed if you attempt to store more tests than are available DO ABS SET TEST 212 is printed when the factory set absorbance gain calibration has been lost from non volatile memory In...

Страница 40: ...e from your instrument supplier This is because in vertical photometry the fill volume and degree of meniscus determine the pathlength and absorbance is proportional to pathlength For lab test results...

Страница 41: ...ires only minimal adjustment from the keyboard to remain calibrated Note that the number entered in test 212 is multiplied by the current absorbance factor set to 1 when reset Thus running test 212 an...

Страница 42: ...ive foam This reader is essentially a maintenance free instrument To ensure maximum trouble free operation the instrument need only be kept dry Although the filters are well sealed their life may be d...

Страница 43: ...on the internal printer After each well is read the result will remain on the display to permit the user to note the result Press ENTER to advance the strip to the next well Problem The lamp does not...

Страница 44: ...oblem The printing is dim or incomplete Solution Use only the specified paper available from your dealer Problem The normal operation of the instrument is suddenly interrupted Solution As with all mic...

Страница 45: ...rference filter temperature and humidity recommendations adapted from Melles Griot Optical Guide 3 1985 pp 266 269 2 Other engineering data supplied by Awareness Technology Inc Palm City FL 1987 2006...

Страница 46: ...42 Stat Fax 303 Plus Operator s Manual Rev J Awareness Technology Inc Notes...

Страница 47: ...rument Appendix 2 is a sheet which may be used to document your stored menu tests Appendix 3 is a set of sample printouts of data from each calculation mode Some modes have been set up to show various...

Страница 48: ...d water wait mix and read Using predispensed dyes reduces the degree of pipetting precision required to reconstitute the check set strips Since the strip reader looks through the sample from top to bo...

Страница 49: ...Awareness Technology Inc Stat Fax 303 Plus Operator s Manual Rev J 45 4 2 Appendix 2 Test Log Microstrip Reader SN ________ User Test Menu Log Page ___ Test Analyte Stored By Date Notes...

Страница 50: ...46 Stat Fax 303 Plus Operator s Manual Rev J Awareness Technology Inc 4 3 Appendix 3 Sample Printouts...

Страница 51: ...Awareness Technology Inc Stat Fax 303 Plus Operator s Manual Rev J 47...

Страница 52: ...he problem support staff at Awareness Technology is happy to assist you and can be reached in the United States by Phone 772 283 6540 Fax 772 283 8020 e mail support awaretech com Mailing address Awar...

Отзывы: