DiagCORE Respiratory Panel 2 Instructions for Use 41
In case seasonal Influenza A H3 strain is detected by the DiagCORE
®
Respiratory Panel 2
assay, two signals will be generated, one for Influenza A and a second one for H3 strain
and shown on the DiagCORE
®
Analyzer screen.
In case a pandemic Influenza A/H1N1/2009 strain is detected, two signals will be
generated, one for Influenza A and a second one for H1N1/2009 and shown on the
DiagCORE
®
Analyzer screen.
For every other pathogen that can be detected with the DiagCORE
®
Respiratory Panel 2, only one
signal will be generated if the pathogen is present in the sample.
8.3
Internal Control interpretation
Controls result
Explanation
Action
Passed
The internal control amplified
successfully.
The run was completed with success,
and all results are validated and can be
reported. Detected pathogens are
reported as positive, and not detected
pathogens are reported as negative.
Failed
The internal control failed.
Positively detected pathogen(s) are
reported, but all negative results (tested
but not detected pathogen(s)) are
invalid.
Repeat the testing using a new
DiagCORE
®
Respiratory
Panel
2
Cartridge.