DiagCORE Respiratory Panel 2 Instructions for Use 19
After sample is introduced inside the DiagCORE
®
Respiratory Panel 2 cartridge, cartridge must
be loaded in the DiagCORE
®
Analyser within 90 minutes.
5.3
Starting the DiagCORE
®
Analyzer
1.
Press the On/Off button on the front of the Analyzer to start the unit.
NOTE
The power switch at the back of the Analytical Module must be set in the “I”
position. The DiagCORE
®
Analyzer status indicators turn blue in this case.
2.
Wait until the Main screen appears and the DiagCORE
®
Analyzer status indicators turn green
and stop blinking.
3.
Log in to the DiagCORE
®
Analyzer by entering the user name and password.
NOTE
The Login screen will appear if User Access Control is activated. If the User
Access Control is disabled, no user name/password will be required and the
Main screen will appear.
4.
In case the Assay Definition File software has not been installed in the Analyzer follow the
instructions prior to running the test (See Section 9 of this document for additional information)
5.4
Running a test
1.
Press the Run Test button at the top right corner of the touchscreen of the DiagCORE
®
Analyzer.
2.
When prompted, scan the sample ID barcode on the nasopharyngeal swab, or scan the
specimen information barcode placed on the top of the DiagCORE
®
cartridge, using the front
barcode reader that is integrated into the DiagCORE
®
Analyzer See figure below).