Starrett FORM 966 Скачать руководство пользователя страница 8

8      FORM 966

ENGLISH

6 & 12

5

4 & 8

1 & 11

3, 9 & 10

2 & 7

BUTTON FUNCTIONS

LOCK, RES, LIMITS and PRESET, 
printed in yellow, are used in 
conjunction with the SHIFT button. 

To engage press the SHIFT button 
first. The SET icon will appear on 
the bottom left corner. Then press 
the button for the desired function.

1 ON/OFF

Power button. Press and hold for 4 seconds.

2 ±

Plus/Minus sets the direction of the reading

3 ZERO

Half a second after releasing the ZERO button the display will 
zero. Do not move from zero during that half second.

4 IN/mm

Toggles the display between English or Metric modes.

5 SHIFT/SET

Dual function. To enable the Lock, Resolution Selection, Preset or  
Limits (Go/No Go) function. When enabled, SET icon will show in 
the bottom left corner of the display.

6 LOCK

Prevents operator modification to the Settings. Press SHIFT/SET 
button, then press and hold the LOCK button (after setting the 
other functions).

7 RES

Resolution. First select the Unit (English/Metric). Press SHIFT/SET 
then the RES (Plus/Minus) buttons. Press the RES button to scroll 
through choices and press the SHIFT/SET to make the selection.

8 LIMITS

Sets the Min and Max values for the Go/No Go function.

9 PRESET

Press the SHIFT/SET button then press the PRESET button to 
enable the PRESET function.

10 ZERO/ABS

Dual Function. Press and hold for two seconds to activate the ABS 
mode or to exit the ABS mode. To zero the display, see #3 above.

11 HOLD

Press HOLD momentarily to capture a reading. The captured reading 
will flash on and off. To disable, press the HOLD button again.

12 MIN/MAX/TIR

Displays the minimum or maximum values captured during 
operation. The TIR function displays the difference between the 
captured readings. 

Содержание FORM 966

Страница 1: ...Trust Is In The Name User Manual FORM 966 ...

Страница 2: ...TO LEA ATENTAMENTE ESTE MANUAL LIRE CE MANUEL AVANT D UTILISER L INSTRUMENT LEIA ATENTAMENTE ESTE MANUAL ANTES DE UTILIZAR O INSTRUMENTO 使用仪器前请阅读本操作手册 DIESES HANDBUCH VOR DER VERWENDUNG DES MESSGERÄTS LESEN LEGGERE ATTENTAMENTE QUESTO MANUALE PRIMA DI UTILIZZARE QUESTO STRUMENTO 01 18 Q ...

Страница 3: ...2900 ELECTRONIC INDICATOR USER MANUAL ...

Страница 4: ...eature Level and Resolution 9 Resolution Selection 9 Setting the Preset Function 10 Setting MIN MAX Limits Go No Go 10 Displaying MIN MAX TIR Values 12 Lock Mode 12 Data Output Connection 13 Battery Installation 13 Appendix A Technical Specifications 14 Appendix B Serial Output Connector 14 Available with Standard and Advanced Features Available only with Advanced Features ...

Страница 5: ...RONIC INDICATOR THIS IS A STARRETT USER GUIDE FOR THE 2900 ELECTRONIC INDICATOR ALL SPECIFICATIONS IN THIS DOCUMENT ARE CORRECT AT TIME OF PRODUCTION AND ARE SUBJECT TO CHANGE PLEASE CONTACT STARRETT FOR FURTHER INFORMATION ...

Страница 6: ...Do not apply lubricant to the spindle or use solvents 9 Do not disassemble or modify the indicator OPERATING INSTRUCTIONS 1 Install batteries included with indicator 2 Lightly clean the contact point 3 Secure the indicator to an appropriate holding device 4 Press the ON OFF button to turn the indicator on 5 Select the unit of measure inch or millimeter by pressing the IN mm button Not applicable t...

Страница 7: ...ENGLISH Stem Cap Data Cable Spindle Bellows Spindle Contact Point Stem Data Output Cover Battery Drawer Back Lug 1 21 32 42 2mm 1 4 6 5mm 1 4 6 5mm 2 5 64 53mm 3 8 9 5mm INDICATOR COMPONENTS DIMENSIONS AND LAYOUT ...

Страница 8: ...the display 6 LOCK Prevents operator modification to the Settings Press SHIFT SET button then press and hold the LOCK button after setting the other functions 7 RES Resolution First select the Unit English Metric Press SHIFT SET then the RES Plus Minus buttons Press the RES button to scroll through choices and press the SHIFT SET to make the selection 8 LIMITS Sets the Min and Max values for the G...

Страница 9: ...0 2 09981 Basic English Metric 0001 001mm 5 12mm 2900 4 09983 Standard English Metric with Selectable Resolution 0005 0001 00005 01mm 001mm 5 12mm 2900 6 09985 Advanced English Metric with Selectable Resolution 0005 0001 00005 01mm 001mm 5 12mm 2900 1M 09986 Basic Metric 001mm 12mm 2900 4M 09988 Standard Metric with Selectable Resolution 01mm 001mm 12mm 2900 6M 09990 Advanced Metric with Selectabl...

Страница 10: ...T button and repeat step 3 Repeat steps 2 and 3 until all the digits are SET 5 Tosetanegativevalue pressPRESETwhentheplus minussigniconisflashing 6 To exit the PRESET function press the PRESET button when the PRESET icon is flashing The SET value will remain displayed 7 Press and hold the ZERO ABS button to return to the active measuring mode SETTINGMIN MAXLIMITS GO NOGOFUNCTION 1 Select the units...

Страница 11: ...eyond the set limits the display will flash and the MIN or MAX LIMIT icon will indicate the direction the limit have been exceeded To exit Press the LIMITS button Press SHIFT SET BUTTON Then LIMITS Button to enter Min Limit Mode Press SHIFT SET BUTTON The LIMIT icon will flash on off Press SHIFT SET BUTTON to increment the value of the digit or the plus minus sign Then press SHIFT SET BUTTON To ca...

Страница 12: ... MIN MAX TIR settings These values are stored in memory Then lock these settings as follows 1 Press the SHIFT SET then press and hold the LOCK button 2 The Lock icon will turn on and stay on see right Once the lock mode has been activated the operator will not be able to change ABS Normal Mode Limits Units PRESET or MAX MIN TIR but will have access to the zero function and be able to view MIN MAX ...

Страница 13: ...e plug as shown below When not using output replace the data cover to protect the electronics BATTERY REPLACEMENT Remove the two Phillips screws that secure the battery drawer as shown below Remove the battery drawer and discard the old batteries Place two new CR2032 batteries into the tray with the positive side facing up Slide the tray carefully into the indicator and secure with the two screws ...

Страница 14: ... displacement stored MX Displacement is minimum peak displacement stored TR Displacement is the T I R stored LL Displacement is the lower or minimum limit setting LU Displacement is the upper or maximum limit setting HL Displacement is a held displacement 1 Sign character minus but no plus 1 Space character IN mode only 6 Digits 1 Decimal point 2 Spaces 1 Character strings to indicate units IN or ...

Страница 15: ...l Order Only State Change High to Low SERIAL INPUT VOLTAGE LEVEL SPECIFICATION Logic 0 Vin 1 30 Volts Logic 1 Vin 1 98 Volts SERIAL OUTPUT VOLTAGE LEVEL SPECIFICATION MINIMUM MAXIMUM Low Level Vss Vss 0 25 1 5mA Load High Level Vdd 0 25 1 25 mA Load Vdd Note Vss 0 Volts Vdd 2 7 to 3 6 Volts ENVIRONMENTAL CONSIDERATION Temperature 10 30 C Humidity 30 85 RH no condensation Atmosphere Non corrosive N...

Страница 16: ...Starrett com ...

Отзывы: