background image

Handbook for the SX remote control Spectrograph    Issue 1, 17/8/20 

 

Methane absorption lines in the near-infrared spectrum of Saturn 

 

 

Hydrogen line doubling from the orbital Doppler shifts in Beta Lyrae 

 
Because the spectrograph uses a concave grating at a steep angle, there is a small amount of residual astigmatism in the 
spectral image. Although astigmatism is normally undesirable, it can be quite useful for expanding the height of the spec-
trum from point sources. This makes the emission and absorption lines more readily visible, without the need to de-focus 
the star image at the slit. As the astigmatism is somewhat variable with wavelength, the spectral height is at a maximum in 
the blue and steadily reduces towards the infrared. This is clearly visible in the spectrum of Alpha Corona Borealis above. 
For most practical purposes, this is of little importance, but many image processing programs include geometric correction 
options that can be used to make the spectrum more uniform. A horizontal perspective correction in ‘Paint Shop Pro’ was 
used to correct the Methane spectrum above. 
 
 
 
 

Advanced adjustments for experienced users: 
 

The height and resolution of the spectrum are both strongly affected by the angle of the small flat mirror in the spectrograph. 
We try to optimise this setting to give best results, but you might want to experiment with the angle to achieve a different 
spectral width or line shape. The theoretical ‘best angle’ is marked by a groove in the floor of the spectrograph, but a varia-
tion of two or three degrees in either direction, will change the spectrum in subtle ways. To make a change, slightly loosen 
the countersunk M4 screw that is visible on the underside of the spectrograph. You can now slew the mirror holder slightly 
to change its angle and then observe the results. Note that the focus of the spectrum will also change as the mirror is twisted, 
so be careful to check that the camera is accurately in focus when checking the result of your adjustment. Multiple changes 
to the spectrum will be seen, as follows: (1) The apparent height and taper of the spectrum (2) The curvature of the emission 
and absorption lines (3) The definition of the lines (4) The start and end positions of the overall spectrum (5) The position of 
the focal plane of the spectrograph. Clearly, you need to thoroughly check that all of these parameters are acceptable, before 
re-tightening the mirror fixing screw. 
 

Spectrograph specifications: 

 

Basic design: 

   

 
Self-collimating concave reflective grating spectrograph with a highly corrected flat field toroidal grating. Minimal attenua-
tion of near UV, due to mostly reflecting optics. Built-in guide camera and calibration source. 
 

Grating specifications: 

 
Groove pitch - 550 grooves per mm at centre of grating. 
Blaze wavelength - 400 nM. 
Spectral efficiency at 400 nM - Greater than 50%. 
Useful spectral range - 340 to 900 nM 
Useful grating aperture - 26 x 26 mm. 
 

 
 

Содержание SX Spectrograph PRO

Страница 1: ...automated most of the spectrograph functions The remote spectrograph allows you to change slits adjust the imaging camera focus and wavelength setting and switch on a calibration lamp or flat field l...

Страница 2: ...es a Lodestar X2 guide camera which observes the image field via a 10 90 beam splitter cube The slit is not directly visible in the field but its position co ordinates are provided and marked by a cur...

Страница 3: ...the wavelengths of the lines from any external source A typical calibration spectrum imaged with an M25C colour camera A spectrum with wavelength markers is provided in the appendix The optical input...

Страница 4: ...d of the cross slide This function is pro vided in the Spectrograph control software As the grating is designed for a compact optical system the available back focal distance for the imaging camera is...

Страница 5: ...Handbook for the SX remote control Spectrograph Issue 1 17 8 20 A general view of the optical and mechanical layout...

Страница 6: ...bottom info strip The spectrograph is automatically started with the slit set to the highest resolution position Number 1 20 microns wide x 1 mm high This is usually the best setting for general purpo...

Страница 7: ...have the Lodestar drivers and software installed please follow these instructions To install the Lodestar driver use the supplied stick to copy the appropriate driver files to a convenient folder on y...

Страница 8: ...size of the spectrograph slit assuming that the 1 mm x 20 uM slit is in use The slit cursor position will probably need adjusting to match your spectrograph The nominal setting will be provided on a l...

Страница 9: ...ee several stars when using a 1 or 2 second exposure and be able to refine the focus until the stars are sharply defined If you have a test target in mind slew to this object and find a good bright gu...

Страница 10: ...ry slew ing the scope slightly East or West before taking another image The Adjust lock position in the Tools menu of PHD2 is good for this If the brightness falls off for both directions then the sli...

Страница 11: ...e such as RSpec see below These simple instructions are sufficient for capturing a spectrum but you will probably want to identify spectral lines and perhaps perform further analysis of the data To th...

Страница 12: ...the calibration chart The small double peak at the right is caused by sodium light pollution and is not in the calibration chart You can see that the spectrum is quite linear and matches the referenc...

Страница 13: ...lected slit The slit widths available 1 20uM x 1mm 2 30uM x 2mm 3 55uM x 2mm 4 115uM x 2mm 5 325uM x 2mm 6 3mm diameter clear aperture For maximum light input with point sources stars or small nebulae...

Страница 14: ...with the angle to achieve a different spectral width or line shape The theoretical best angle is marked by a groove in the floor of the spectrograph but a varia tion of two or three degrees in either...

Страница 15: ...2 mm Position 5 325 uM x 2 mm Position 6 3mm x 3 mm circular Input Output Female T2 thread 42 x 0 75 mm input Male T2 thread output Input back focal distance approximately 37 mm Output back focal dist...

Страница 16: ...1 The equipment shall only be used for normal purposes described in the standard operating instructions and within the relevant safety standards of the country where the equipment is used 2 Repairs u...

Отзывы: