18
Information and settings
S.Box Duo S and S.Box Duo ST
4. Settings menu
To access the settings, tap the touch key
when the device is in standby mode. In the following
sequence of settings:
Tapping on
or
allows you to decrease or increase the value of the parameter or to
deactivate or activate the displayed function,
Tapping on
allows you to access the next display,
maintaining the ramp button
pressed, then tapping simultaneously the touch key
allows
coming back to the previous display.
The display turns off automatically after two minutes without pressing a key or by tapping
.
Views of the display
:
❶
Tap the touch key
in standby mode.
Ramp time: time taken by the device to reach
the prescribed pressure.
This parameter is displayed if the ramp has
been enabled by your Home Care Provider.
By tapping the key
or
as many times
as necessary, you can decrease or increase the
value displayed or disable the ramp feature in
selecting
.
Possible settings: OFF and from 05 min to the
ramp time set by your Home Care Provider.
❷
Tap the touch key
Comfort pressure: pressure level produced when
the ramp starts up.
This parameter is displayed if the ramp time is
≠ OFF and if your Home Care Provider did not
lock your access to this setting.
You can decrease or increase the value
displayed by tapping the key
or
repeatedly.
Possible settings:
from 4 cm H
2
O to the set pressure in CPAP
mode.
from 3 cm H
2
O to the expiratory pressure in
BPAP (bilevel) mode.
❸
Tap the touch key
IS (Intelligent Start)
This function is displayed if it has been activated
by your Home Care Provider and if he did not
lock your access to this setting.
You can deactivate (OFF) or activate (ON) this
function by tapping the key
or
.
Possible settings: IS: ON and IS: OFF.