Humidity Alarm Settings:
Local Alarm: The Local Alarm includes 3 components; Buzzer,
Alert LED and the on-screen Bell icon. These are all triggered if
the configured temperature values are out of range. By default
toggling the button.
Note: The buzzer will continue to sound unless muted on the tag
itself.
Min & Max Humidity Thresholds: Set the minimum and
be triggered when a value is out of its configured range.
Alarm Delay Period: Select
must be out of a configured range before an alarm is triggered.
For example; if the alarm delay period is set to 5 minutes and the
lue is out of its configured range, the alarm will only be
triggered if its value does not normalize within the set time of 5
minutes.
Note: The tag will immediately move out of alarm mode when
the humidity is back within the configured limits.
7.
Apply the tag configuration by tapping the check icon
.
8.
Tap Apply to apply the settings to the tag(s).
9.
Ensure all tag configurations are successful and then tap Done.