14  

 

   

Installation and Use 

Test 

Run a few normal fasteners. The ePro will shut off the tool. Let go of the throttle (the ePro will reset the internal shut-off valve and 
the tool will turn back on). Check the torque results.  

The system is NOT defaulted for REHIT error detection, i.e. re-hitting an already tightened fastener.  This feature must be set up 
through MIN PULSE COUNT in the ADMIN MENU. Please see Administration Functions to set the system up to detect a REHIT fastener 
and error or display as a fault. 

Run the tool in free air and let go of the throttle and the program should ignore the run, display LOW TQ or ABORT. Slip off or strip 
out a fastener and the display should ignore, error or read SLIP. 

If the results obtained are not correct, press ENTER and re-program the application or you can go to the ADMINISTRATION functions
by depressing the 

hidden

 button under the ePro logo. Use the BACK key to return. 

Administration Functions 

From the RUN screen, access the Administration functions via the button located under the ePro logo. By quickly pressing the ePro 
logo, the following options will display. Scroll the functions using the UP and DOWN buttons. 

NOTE

: After adjusting function(s), press 

ENTER to accept the change, then press the BACK UP key until the run screen appears. 

CAL CORRECT: Provides an easy transducer calibration adjustment to an independent system or torque wrench. When using a torque 
transducer connected to the controller, make a rundown and compare the torque reading from the controller to an independent 
system. The torque reading from the last rundown appears in the top line of the display as TQ:. By scrolling up or down, enter the
value from the independent measurement in the bottom line of the display as NEW TQ:. This changes the transducer calibration value
by the percent difference of the two. Future readings should agree with the independent system.  

START PSI (ADJUST): This screen allows changes to the tool’s starting PSI level (air pressure). This is called START PSI because the 
ePro default setting will RAMP up the PSI from this “start” value in order to reach TARGET TQ as quickly as possible.  

PRE-TRIGGER: This allows an early shut-off of the tool in order to control any torque “over-shoot” that may occur in certain difficult
applications. Press the ENTER button and two values will appear. The “A” value in the amount of  “over-torque” that occurred from
the last rundown. The “PT” value is the amount of pre-trigger torque you wish to initiate. Scroll UP to set the amount of pre-trigger
desired, then press ENTER to activate this feature. 

THRESHOLD: This is the torque setting where the ePro deems that a fastening event has begun. This is automatically set to 30% of
the TARGET TQ or another pre-determined value. Scroll to alter. Press ENTER to activate change. 

PULSE DELAY: The ePro can ignore the initial pulses from the tool. This is required to eliminate any pulse “spiking” from the initial
pulse(s). This can happen due to the high free-speed of the tool. This is defaulted to (5) pulses to cover “worst case” applications.

JOINT CONDITION: This feature adds extra pulses to the end of the fastening cycle. This is helpful for applications that may “relax”
or lose torque after fastening. PSI is the pressure at which the conditioning will occur. START is the START PSI setting of the ePro. 
Scroll to the PSI desired for the conditioning. Press ENTER. PULSES will appear. Scroll to set the number of conditioning PULSES. 
Press ENTER to activate. 

DOWNSHIFT: This allows the end of the fastening cycle to be slowed down in order to control torque “over-shoot”. TORQUE will 
appear when ENTER is pressed showing the TARGET TQ. Scroll down to set the TORQUE at which to initiate DOWNSHIFT. Press 
ENTER. PSI will appear. Scroll to the PSI level which the tool will run until TARGET TQ is achieved. 

MAX PULSE COUNT: This sets a maximum allowable PULSE COUNT before the cycle is REJECTED. If a strip or cross-thread occurs, 
this can help detect the bad cycle. It can also detect a failing pulse pack unit. Press ENTER and LAST RUN and MAX PULSE CT setting
will appear. The setting can be tested at this point. Watch the screen and run some good and bad cycles. Adjust MAX PULSE CT 
setting until satisfied. Press ENTER to activate.

MIN PULSE COUNT: This sets a minimum allowable PULSE COUNT before the cycle is REJECTED. If a previous tightened fastener or a 
cross-thread occurs, this can detect the bad cycle. Press ENTER and pulse counts from the LAST RUN and the MIN PULSE CT setting
will appear. Scroll UP or DOWN to set the MIN PULSE CT. The setting can be tested at this point. Watch the screen and run some 
good and bad cycles. Adjust MIN PULSE CT setting until satisfied. Press ENTER to activate. 

작업 스크린으로부터

, PSI-C

로고의 아래에 위치하는 단추를 통해

 Administration 

기능들에 접근 하세요

. PSI-C

로고를 빨리 누

르는 것에 의해,아래의 옵션들은 디스플레이 될 것 입니다

. UP

키와

 DOWN 

키를 사용하여 기능들을 스크롤 하세요

.NOTE:

Function(s)

를 조정하고 난 후에

,

변경값을 조정하기 위해

 ENTER

키를 누른 후 운용스크린이 나타날 때까지

 BACK UP

키를 누르

세요

.

CAL CORRECT:

독립 시스템 또는 토크 렌치에 간단한 트랜스듀서 칼리브레이션교 조정을 제공 합니다.토크 트랜스듀서와

연결된 콘트롤러를 사용 할 때

,

콘트롤러에서 독립된의 시스템까지

 Rundown Data

를 만들고 토크를 비교 합니다

.

마지막

Rundown Data

로부터 측정된 토크값은

TQ

로 디스플레이에 첫번째 라인에 나타 납니다

: UP

키 또는

 DOWN

키를 스크롤하여 새

로운 값을 독립된 장치로부터 디스프레이의 맨 밑줄에

 NEW TQ

를 입력합니다

. :

이것은 둘의 퍼센트 차이에 의해 트랜스듀서

칼리브레이션값을 바꿉니다

.

향후의 측정값은 독립의 시스템과 일치해야 합니다

.

START PSI(ADJUST):

이 스크린은

 tool

의 시작

 PSI 

레벨

(

공압

)

의 변경을 허용합니다

.

이것은

 START PSI

라고 불리 우는데

 PSI-C

초기 설정은 가능한한 빨리

 "start" 

값에서 타겟 토크에 도달하기 위하여

 RAMP

를 설정 합니다. 

PRE-TRIGGER:

이것은 어려운 작업에서 발생 될지도 모르는

 "over-shoot" 

토크를 컨트롤 하기 위해서 빠르게 툴의 정지를 허용

합니다

. ENTER

키를 누르세요

.

그러면

 2

개의 값이 나타날 것 입니다

.

마지막 런다운에서 일어났던

“over-torque”

의 양이

 “A”

입니다

. “PT”

값은 시작하고 싶은 프리 트리거 토크의 힘의 총계입니다

.

요구되는 프리 트리거 총계를 설정하기 위해

 UP

키를

누르고 이 특징을 활성화시키려면

ENTER

키를 누르세요

.

THRESHHOLD:

이것은

 PSI-C

가 조임 과정이 시작되었다라고 간주되는 곳의 토크 셋팅입니다

.

이것은 자동적으로

 TARGET TQ

30%

또는 미리 정해 놓은 다른 값에 조정 됩니다

.

스크롤하여 바꾸세요

.

다음단계로 변경하려면

 ENTER

키를 누르세요

.

PULSE DELAY: PSI-C

는 툴로부터 최초의 펄스를 무시 할 수 있습니다

.

이것은 최초의 펄스

(s)

로부터 어떤 펄스

“spiking”

라도 제

거하는 것을 요구 합니다

.

이것은 도구의 높은 프리 스피드 때문에 일어날 수 있습니다

.

이것은

 “

최악의 경우

작업들을 보호

하기 위해

 (5)

개의 펄스들이 실행되지 않습니다

.

JOINT CONDITION:

이 특징은 조임 사이클의 마지막에 여분의 펄스를 추가합니다

.

이것은 잠기고 난 후에

 “relax" 

하거나 조임

후에 토크를 잃을지도 모르는 공정에 도움이 됩니다

. PSI

는 조절 상태의 압력입니다

. START

 PSI-C

 START PSI

설정입니

다.

PSI

를 스크롤하여 상태를 조정하세요

. ENTER

키를 누르세요

. PULSES

가 나타날 것 입니다

. PULSES

상태의 넘버를 셋팅

하기 위해 스크롤 하세요

.

다음 단계로 가려면

 ENTER

키를 누르세요

.

DOWNSHIFT:

이것은 조임 사이클의 마지막에 토크를 제어하기 위하여

 “over-shoot” 

을 느리게 하는 것을 허용 합니다

.

TORQUE

 ENTER

키가 눌러지면

 TARGET TQ

가 나타나게 될 것입니다

. DOWNSHIFT

를 시작하는

 TORQUE

를 설정하기 위해

DOWN

키를 스크롤 하세요

. ENTER

키를 누르세요

. PSI

가 나타날 것 입니다

. TARGET TQ

가 나올 때까지 툴을 운용하고

 PSI

레벨

을 스크롤 하세요

.

MAX PULSE COUNT:

이것은 사이클이 거부되기 전에 최대 허용 할 수 있는

 PULSE COUNT

를 설정합니다

.

만일

 strip 

또는

 cross-

thread

발생되면

,

이것은 불완전한 사이클을 찾는 것을 도울 수 있습니다

.

이것은 또한 결함이 있는 펄스 패키지 유닛을 찾을

수 있습니다

. ENTER

키를 누르면 마지막 공정과 최대

 PULSE CT

셋팅을 나타날 것 입니다

.

셋팅은 이 시점에서 테스트 되어 질

수 있습니다

.

스크린을 보면서 유효한 사이클과 불완전한 사이클을 운용하세요

.

만족 할 때까지

 MAX PULSE CT

설정을 조정

하세요

.

다음 단계로 가려면

ENTER

키를 누르세요

.

MIN PULSE CONUNT:

이것은 사이클의이 거부되기 전에 최소 허용 할 수 있는

 PULSE COUNT

를 설정합니다

.

만일 이전의 확실

히 조여진 조임체 또는

 cross-thread

가 발생되면

,

이것은 불완전한 사이클을 찾는 것을 도울 수 있습니다

. ENTER

키를 누르면

마지막 공정과 최소

 PULSE CT

셋팅이 나타날 것 입니다

. UP

키와

 DOWN

키를 스크롤하여 최소

 PULSE CT

를 설정 하세요

.

셋팅

은 이 시점에서 테스트 되어 질 수 있습니다

.

스크린을 보면서 완전한 사이클과 불완전한 사이클을 운용하세요

.

만족 할 때까

 MIN PULSE CT

설정을 조정 하세요

.

다음 단계로 가려면

 ENTER

키를 누르세요

.

Administration Functions

13

PSI-C

몇몇개의 일반적인 조임체를 조이세요

. PSI-C

는 툴을 정지시킬 것 입니다

.

레버를 놓으세요

.(PSI-C

는 내부의 셧 오프 밸브를 재

셋팅하고 툴을 되돌려 놓을 것 입니다

. )

토크값을 점검하세요

.

시스템은

 REHIT 

에러 발견을 위해 초기화 되지 않을 것 입니다

.

re-hitting

은 이미 조임체가 조여짐

.

이 특징은

 ADMIN MENU

에서

 MIN PULSE COUNT

를 통해 설정되 있어야 합니다

. REHIT fastener

와 에러 발견 또는 오류로 인한 표시 등의 시스템 셋업은

Administration

기능을 참고 하십시요

.

.

Free air

에서 툴을 운용하고 레버를 놓으면 프로그램은 툴 운용을 무시하게 되고

LOW TQ

또는

 ABORT

를 나타낼 것 입니다

.

Slip off

또는

 Strip out

된 조임체는 디스플레이에 무시되고 에러가 되며 또는

 SLIP

으로 나타 납니다

.

만일 얻어지는 결과들이 정확하지 않으면

 ,ENTER

를 누르고 재프로그램 하거나 그렇지 않으면 아래에서 숨겨졌던 단추를 눌

 ADMINISTRATION 

기능에 들에 갈 수 있습니다.되돌아 가려면

 BACK 

키를 사용 하세요

.

Содержание PSI-C

Страница 1: ...the equipment or described in this manual Stanley Assembly Technologies Applications Productivity Center 5800 Enterprise Court Warren MI 48092 3476 Tel 586 393 1100 Fax 586 393 1200 Toll Free in the US 1 877 787 7830 www stanleyworks com PSI C 0406 THE STANLEY WORKS ALL RIGHTS RESERVED G G G G G G G G G G G G G G G G G G rUzUlUjG 㟬䏭ⷸ㐐GἼ㷐ẠGᴴ G _XTXXG 䊀䆠 G_㵜G W 䝬G ࣜࣜࣜऌऄऋऊँࣶࣜࣜ ࣲࣩࣩࣴࣳࣳ ࣳࣤ堆ࣥࣜࣨࣜंࣽऔࣶࣜࣜ ࣩࣴ...

Страница 2: ... Air Supply Requirement 6 Tool Compatibility and Transducer Pulse Tools 7 Layout and Operator Interface 8 Main Menu and Application Programming 9 Parameter Set Up 10 Test 11 Administration Functions 12 Error Proofing OK NOK 13 Normal Operation Run Mode 13 Inputs Outputs I O 14 RJ45 15 Networking 15 Service and Maintenance 15 Troubleshooting and Adjustments 16 2 PSI C ...

Страница 3: ...Only allow suitably qualified personnel to install program or maintain this equipment and or system These persons must be knowledgeable of any potential sources of danger and maintenance measures as set out in the Installation Operations and Maintenance manual This product must be transported stored and installed as intended and maintained and operated with care to ensure that the product function...

Страница 4: ...he standard configuration utilizes a CE approved power supply that is UL listed It consists of a wall mount 120VAC transformer with a 16VAC 50 Hz at 1 Amp power output Power supplies for other International standards Japan 110 v 50 Hz 1 Amp part 10565 Australia 240 v 50 Hz 1 Amp part 10566 United Kingdom 240 v 50 Hz 1 Amp part 10567 Europe 220 v 50 Hz 1 Amp part 10568 To Avoid Injury Plug power co...

Страница 5: ... rated up to 7 amps at 48V AC DC Printer optional Internal Thermal Printer for documentation of cycle data onto individual labels 2 x 4 adhesive backed Serial Port RS 232 port RJ45 connector Interface Dual Line 0 5 LCD backlit 16 character Five button keypad Up Down Enter Menu Back ADMIN Back button Momentary for Batch reset hold down 2 sec for Over ride LED Indicators CYCLE OK NOK Indicators BATC...

Страница 6: ...valve and will not allow air flow back thru the lubricator This type is not recommended for use with the ePro FSC The tool air hose cable for the EITP EITM and ESPT series of tools is called the GEMINI The GEMINI air hose includes an integral transducer cable and is available in 3 8 I D coiled or straight and available in many lengths A separate transducer cable is also available Prevent any unnec...

Страница 7: ...he torque required If prevailing torque is present then an over sized tool may be required If you desire to utilize any other brand of standard pulse tool Uryu Atlas Copco etc contact your salesperson for a transducer attachment for your tool This allows use of these common tools to be controlled with the PSI C process controller wzpTjG䆮䔎 㦖G G㭒㣪䞲Gぢ ⰢG㞚 G㓺䌶ⰂGwlG䔎 㓺 G䗚㓺䒊ὒG 䢎Gᾦ䢮 㦚GṖ㰧 UG 䞲Gộ㦖G 㔶㦮G 㧒㯞ⰾ...

Страница 8: ...45 REMOTE KEYPAD LIGHTS etc 12 24V I O INPUTS OUTPUTS POWER SOCKET ePro URYU option POWER SOCKET STD 3 8 NPT AIR OUTPUT GEMINI HOSE CONNECT I O SCHEMATIC DATA X FER NETWORK SERIAL ETHERNET or DCX PFCS 1 2 NPT AIR INPUT TRANSDUCER CABLE CONNECT GEMINI HOSE PSI C ...

Страница 9: ...Installation and Use 9 11 10 20 23 22 4 7 5 19 6 17 9 8 18 13 15 24 27 30 32 1 21 3 34 35 12 33 25 29 28 26 2 31 16 36 37 38 ePro FSC Controller Assembly Assembly Part No 11024 PSI C PSI C ...

Страница 10: ...creen Press the ENTER key and the ePro automatically checks and sets the ZERO and MAX values of the transducer connected TARGET TORQUE Enter the target torque of the application by scrolling up or down press ENTER HIGH LIMIT Scroll to set the upper torque specification press ENTER LOW LIMIT Scroll to set the low limit torque specification press ENTER BATCH COUNT Scroll up to set the number of fast...

Страница 11: ...LETE CAL TRANSDUCER CHECKING CAL CAL TRANSDUCER CHECKING MAX PARAMETER 1 NOT SETUP SET TARGET TORQUE 50 0 NOM CAL VALUE TQ 75 0 CAL TRANSDUCER CHECKING ZERO CAL TRANSDUCER CHECKING MAX CAL TRANSDUCER COMPLETE SET HI LIMIT TORQUE 57 5 Use the UP DWN arrows to find the parameter set desired Press ENTER to select Default HI LO Limits are set to 15 of Torque Target 11 PSI C PSI C ...

Страница 12: ...OPERATION FLOWCHART INITIAL SETUP FSC CONT SET BATCH TIMER SECOND S 4 SET BATCH COUNT COUNT 2 TQ 50 5 PS 2 CT 02 TIMER 004 This is the RUN screen SET LO LIMIT TORQUE 42 5 12 PSI C PSI C CONT ...

Страница 13: ... the tool will run until TARGET TQ is achieved MAX PULSE COUNT This sets a maximum allowable PULSE COUNT before the cycle is REJECTED If a strip or cross thread occurs this can help detect the bad cycle It can also detect a failing pulse pack unit Press ENTER and LAST RUN and MAX PULSE CT setting will appear The setting can be tested at this point Watch the screen and run some good and bad cycles ...

Страница 14: ...usually 50 60 of the specified final torque Press ENTER Note BATCH COUNT must be programmed for correct sequencing The Run Screen will now show PTQ when running the PRETORQUE or snug torque PRETORQUE all fasteners When PRETORQUE BATCH is finished OK The Run Screen will change to show TQ Final Torque all fasteners Note MIN PULSE COUNT may need to be adjusted to accept final torque while still rejec...

Страница 15: ...START 65 START PSI PSI 50 ADMIN MENU START PSI ADMIN MENU DOWNSHIFT DOWNSHIFT TORQUE 50 0 JOINT CONDITION PULSES 0 ADJUST DOWNSHIFT PSI 35 START 50 DYN TQ 0 0 STATIC TQ 0 0 CAL CORRECT CAL 200 0 CONTINUED BELOW PULSE DELAY PULSE S 5 ADMIN MENU PULSE DELAY PRETRIGGER PT 00 0 A 00 0 ADMIN MENU PRETRIGGER ADMIN MENU MAX PULSE COUNT LAST RUN 32 MAX PULSES 200 ADMIN MENU MIN PULSE COUNT LAST RUN 37 MIN...

Страница 16: ...0 UP SET DWN CLR ADMIN MENU CYCLE COUNTER CYCLE COUNTER CYCLES 0 ADMIN MENU BATCH COUNT SET BATCH COUNT COUNT 2 SET BATCH TIMER SECOND S 4 ADMIN MENU BATCH TIMER ADMIN MENU LO LIMIT ADJUST LO LIMIT TORQUE 43 2 ADJUST HI LIMIT TORQUE 52 8 ADMIN MENU HI LIMIT CONTINUED BELOW ADMIN MENU CYCLE DELAY SET CYCLE DELAY DELAY 1000 CONTINUED ON NEXT PAGE 16 PSI C PSI C CONT ...

Страница 17: ...T ADMIN MENU TIME DATE 03 08 07 01 49 UP SET DWN EXIT SET TIME DATE MONTH 03 SET TIME DATE DAY 08 SET TIME DATE YEAR 07 CONTINUED ON NEXT PAGE ADMIN MENU STATS STATS SAMPLES 10 DATA UP CLR DWN DWNLOAD ADMIN MENU DATA DATA 28 800 8 N 1 DATA DATA CLEARED CONTINUED BELOW 17 PSI C PSI C CONT ...

Страница 18: ... TQ 0 0 PRETORQUE THOLD 0 0 PRETORQUE PSI 50 0 ADMIN MENU ADMIN LOCK OFF LEARN ADMIN CODE CODE ADMIN MENU RAMP SPEED MED ADMIN MENU RAMP SPEED HI ADMIN MENU FILTER SET FILTER FREQ FREQ 1150 ADMIN MENU RAMP SPEED LOW UNIT IP ADDRESS 192 ADMIN MENU ETHERNET OPERATION FLOWCHART ADMIN MENU FSC CONT CONTINUED BELOW ADMIN MENU RAMP SPEED OFF 18 PSI C PSI C CONT ...

Страница 19: ...OUT is recommended when only one PSET is being utilized Pressing the BACK button clears the batch The BATCH NOK output fires and turns on the red BATCH LED Pressing and holding the BACK button for 1 2 seconds puts the ePro in OVER RIDE mode OVER RIDE means that the system has full power without any torque control or shut off capabilities The ePro has a two line main display This shows the final TO...

Страница 20: ...of either polarity I O SCHEMATIC OUTPUTS PIN 1 COMMON PIN 2 CYCLE OK PIN 3 CYCLE NOK PIN 4 BATCH OK PIN 5 BATCH NOK INPUTS PIN 6 DISABLE TOOL PIN 7 BATCH REJECT RESET PIN 8 MANUAL WRENCH ACCEPT PIN 9 PSET BIT 3 BINARY PIN 10 PSET BIT 2 PARAMETER PIN 11 PSET BIT 1 SELECT PIN 12 INPUT COMMON I O CONNECTOR PLUG OPTIONAL PART 11424 BATCH DISABLE BATCH BATCH 12 11 10 9 8 7 6 5 4 3 2 1 0 0 1 PS1 0 1 0 P...

Страница 21: ...a Collection and Download If the ePro FSC has the optional DTU Data Transfer Unit part 10771 all parameter sets and rundown data from the last 700 readings per Pset is stored in the memory of the DTU The SigmaSIX Data Transfer Kit part 11556 can be used to download this data to a Microsoft Excel spreadsheet The kit includes the SigmaSIX Data Transfer Software Program and all cables and adapter req...

Страница 22: ...led and the ePro is being converted to PFCS communication follow the instructions below PFCS EtherNET network card Remove jumpers from main control board JP1 and JP2 PFCS Serial network card Remove all jumpers from main control board JP1 and install jumpers onto all positions on JP2 Once the Data Collection Cable Assembly is correctly assembled Open the SigmaSIX Data Collection Software Program Se...