5-13
Remote Programming
Aliasing Effects
In any sampled data stream, it is possible to sample a high frequency signal such that it will appear to be a
much lower frequency. This is called aliasing.
For example, suppose the lock-in is detecting a signal near 1 Hz with a relatively short time constant. The X
output will have a DC component and a 2 Hz component (2xf). If the sample rate is 2 Hz, then the samples
may be taken as illustrated below.
The samples represent a sine wave much slower than 2 Hz that isn't actually present in the output! In this
case, a much higher sampling rate will solve the problem.
Aliasing occurs whenever the output signal being sampled contains signals at frequencies greater
than 1/2 the sample rate. The effect is most noticeable when trying to sample an output frequency at an inte-
ger multiple of the sample rate (as above). The above aliasing problem will be the same for a 1 kHz output
(500 times the sample rate) as for the 2 Hz output.
Generally, the highest possible sample rate should be used given the desired storage time. The lock-in time
constant and filter slope should be chosen to attenuate signals at frequencies higher than 1/2 the sample rate
as much as possible.
SRAT (?) {i}
The SRAT command sets or queries the data sample rate. The parame-
ter i selects the sample rate listed below.
i
quantity
i
quantity
0
62.5 mHz
7
8 Hz
1
125 mHz
8
16 Hz
2
250 mHz
9
32 Hz
3
500 mHz
10
64 Hz
4
1 Hz
11
128 Hz
5
2 Hz
12
256 Hz
6
4 Hz
13
512 Hz
14
Trigger
SEND (?) {i}
The SEND command sets or queries the end of buffer mode. The param-
eter i selects 1 Shot (i=0) or Loop (i=1). If Loop mode is used, make sure
to pause data storage before reading the data to avoid confusion about
which point is the most recent.
TRIG
The TRIG command is the software trigger command. This command
has the same effect as a trigger at the rear panel trigger input.
TSTR (?) {i}
The TSTR command sets or queries the trigger start mode. The parame-
ter i=1 selects trigger starts the scan and i=0 turns the trigger start fea-
ture off.
STRT
The STRT command starts or resumes data storage. STRT is ignored if
storage is already in progress.
1 second
Содержание SR830
Страница 5: ...1 4...
Страница 11: ...SR830 DSP Lock In Amplifier 1 10...
Страница 13: ...2 2 Getting Started...
Страница 17: ...2 6 The Basic Lock in...
Страница 23: ...2 12 Outputs Offsets and Expands...
Страница 25: ...2 14 Storing and Recalling Setups...
Страница 31: ...3 4 SR830 Basics...
Страница 33: ...3 6 SR830 Basics...
Страница 37: ...3 10 SR830 Basics...
Страница 53: ...3 26 SR830 Basics...
Страница 74: ......
Страница 83: ...4 30 Rear Panel...
Страница 107: ...5 24 Remote Programming...
Страница 113: ...5 30 Remote Programming...
Страница 117: ...5 34 Remote Programming...
Страница 121: ...6 4 Performance Tests...
Страница 123: ...6 6 Performance Tests...
Страница 125: ...6 8 Performance Tests...
Страница 129: ...6 12 Performance Tests...
Страница 131: ...6 14 Performance Tests...
Страница 133: ...6 16 Performance Tests...
Страница 139: ...6 22 Performance Tests...
Страница 145: ...7 2 Circuit Description...