3.4
Commands
3 – 19
by
OLSR?
or
*CLS
, the overload status bits will stay cleared even
though the overload condition may persist and remain reported
by
OVLD?
.
OVLD?
Example:
6
implies that the bias is
not
overloaded; the transimpedance stage
(
V
bias
−
i
in
×
R
F
)
is
overloaded; and the output
is
overloaded.
Reference Clock State
RCLK?
Query the current source and lock state of the reference clock
of the preamplifier. The responses are
INTERNAL 0
,
EXTERNAL 1
,
and
UNLOCKED 3
.
This command complements the RCSR status register described in
Section 3.5.9, but there is no one-to-one correspondence between
the response of
RCLK?
and bits in RCSR. Once cleared by
RCSR?
or
*CLS
, the reference clock status bits will stay cleared even though
the reference clock state may persist and remain reported by
RCLK?
.
RCLK?
Example:
UNLOCKED
3.4.8
Interface commands
The Interface commands provide control over the interface between
the SIM918 and the host computer.
Reset
*RST
Reset the SIM918 to its default configuration.
*RST
sets the following:
1. Gain to 10
6
V
/
A.
2. Autozero on.
3. Input connected.
4. Input shield to ground.
5. Bias to ground.
6. Bias shield to ground.
7. Reference clock direction to input.
8. The phase-locked loop to inactive when autozero is o
ff
(
APLL
OFF
).
9. Clock oscillator to stop during idle time (
AWAK
OFF
).
SIM918
Precision Current Preamplifier
Содержание SIM918
Страница 4: ...ii Contents SIM918 Precision Current Preamplifier ...
Страница 7: ...General Information v SIM918 Precision Current Preamplifier ...
Страница 12: ...x General Information SIM918 Precision Current Preamplifier ...
Страница 62: ...3 30 Remote Operation SIM918 Precision Current Preamplifier ...
Страница 63: ...4 Circuit Description In This Chapter 4 1 Schematic Diagrams 4 2 4 1 ...