3 – 18
Remote Operation
Pulse
¬
STATUS Mode
PSTA(?)
{
z
}
Set (query) the Pulse
¬
STATUS mode
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
When
PSTA
ON
is set, all new service requests will only
pulse
the
¬
STATUS signal LOW (for a minimum of 1
µ
s). The default behavior
is to latch
¬
STATUS
LOW until a
*STB?
query is received.
A reset does not alter
PSTA
. The value in boldface above is the
power-on value.
PSTA OFF
Example:
Last Button
LBTN?
Query the number of the last button pressed. The response is
LBTN?
Last button
1
[GAIN ]
2
[GAIN ]
3
[AUTOZERO]
4
[Output 1 pps sync]
5
[INPUT Open]
6
[INPUT Shield]
7
[BIAS GND]
8
[BIAS Shield]
9
Both [GAIN ] and [GAIN ] (autocalibrate)
10
A button held upon power-on (reset)
The value
0
is returned if no button was pressed since the last
LBTN?
.
A query of
LBTN?
always clears the button code, so a subse-
quent
LBTN?
will return
0
.
LBTN?
Example:
5
Overload
OVLD?
Query the current overload condition. The response is
OVLD?
Overloaded
1
Bias
2
Output
4
Bias
+
Output
Combination overloads are reported by summing the values of the in-
dividual overload flags. This command complements the OLSR sta-
tus register described in Section 3.5.7, and the three overload flags
correspond one-to-one with bits in OLSR. However, once cleared
SIM918
Precision Current Preamplifier
Содержание SIM918
Страница 4: ...ii Contents SIM918 Precision Current Preamplifier ...
Страница 7: ...General Information v SIM918 Precision Current Preamplifier ...
Страница 12: ...x General Information SIM918 Precision Current Preamplifier ...
Страница 62: ...3 30 Remote Operation SIM918 Precision Current Preamplifier ...
Страница 63: ...4 Circuit Description In This Chapter 4 1 Schematic Diagrams 4 2 4 1 ...