The  air  supply  system  must  be  provided  with  a  valve  (plug)  for  a  complete  stoppage  of  air 

supply in the case of maintenance and/or resetting of the system, or in cases of major changes in loads 
of the system.  

 

Attention 
 

 

It  is  not  advisable  to  use  compressed  carbon  dioxide  (CO

2

)  tanks  because  rapid  filling  of 

pneumatic supports may result in their icing up.   

 

 

 

 
2.   Mounting of the system 

 

2.1    Assembly of the system  

 

2.1.1  Safety  

 
The  weight  of  the  optical  table  could  reach  100  kg.  For  this  reason,  to  prevent  injuries  during 

assembly of the system,  lifting mechanisms (a loader)  and technological  supports  must be prepared in 
advance.    

 
2.1.2  Assembly of the frame 

 

 Assembly of the frame is shown in fig. 4. 

 

 

Attach the brace into the leg assembly by M10x120 screws through the washers. 

 

 

Screw the solid leveling elements into the leg assembly. 

 

 

 

 

Fig. 4. Assembly of the frame 

Содержание 1VIS10

Страница 1: ...s Lithuania Phone 370 5 2651474 Fax 370 5 2651483 E mail sales standa lt Manufacturer of Opto Mechanical equipment for research industry and education http www standa lt Vibration isolation system 1VIS10 User manual Standa 2018 ...

Страница 2: ...embly of the system 2 1 1 Safety 2 1 2 Assembly of the frame 2 1 3 Connection and installation of pneumatic vibration isolation components 2 1 4 Fitting the collector and the support plates 2 1 5 Assembly of the table 2 2 Preparation of the system for operation and vibration isolation of the table 2 2 1 Assembly of level controls 2 2 2 Connection of air supply pipes 2 2 3 Setting of level controls...

Страница 3: ...ifferent sizes and loads Specification Symbol Value Unit Vertical movement range lz 14 mm Precision of automatic level control Δ lz 0 3 mm Pressure from source of air p 600 6 bar 87 PSI kPa Air supply opening Ø 6 mm Operating humidity max φ 90 Operating temperature range T 10 50 ºC Load bearing capacity at 6 bar per isolator Q 160 kg Fig 1 Specification of 1VIS10 A specification of a pneumatic sup...

Страница 4: ...or research industry and education http www standa lt 4 Fig 2 Level control model 1LV 1WM Optional An overpressure relief valve can be provided 1 Precision of level setting 0 3 mm 2 Materials Body and lever anodised aluminium brass steel plastic A typical vibration isolation system is shown in fig 3 Fig 3 A typical vibration isolation system ...

Страница 5: ...ty of other sources of vibration such as elevators ventilation systems industrial equipment and airflows Warning The system is a metal made current conductor If used together with electrical devices the table must be earthed 1 3 Air supply requirements The operation of the system requires a constant supply of air After the system has been filled in and set air is used only for the operation of lev...

Страница 6: ...t advisable to use compressed carbon dioxide CO2 tanks because rapid filling of pneumatic supports may result in their icing up 2 Mounting of the system 2 1 Assembly of the system 2 1 1 Safety The weight of the optical table could reach 100 kg For this reason to prevent injuries during assembly of the system lifting mechanisms a loader and technological supports must be prepared in advance 2 1 2 A...

Страница 7: ...e collet fittings until tight Pull the pipes accurately to make sure that the connection is secure The length of the pipes between the fittings must be optimal with a small extra length Warning Cut the edges of the pipes using a sharp cutter at a straight angle It is not advisable to use scissors because deformation of the pipe edges may lead to leaks from the system Attach the pneumo isolation el...

Страница 8: ...d education http www standa lt 8 2 1 4 Fitting the collector and support plates Attach the collector to the leg assembly by M5x30 screws Insert the hex key 3HK 5 into the hole of the piston and hold it so that the torque moment doesn t transmit to the diaphragm Attach the support plate to the piston by M6x16 screw Fig 6 Fitting the collector and the support plates ...

Страница 9: ...t Manufacturer of Opto Mechanical equipment for research industry and education http www standa lt 9 2 1 5 Assembly of the table Place the table accurately on top of the pneumatic supports If you need to adjust the location of the table lift the table over the supports Fig 7 Assembly of the table ...

Страница 10: ...arrying capacity applied to the centre of the shape formed by the pneumatic supports Groups of pneumatic supports serve as three supports for the system The aggregate centre of gravity of the optical table and of the useful load present on the table must be within the boundaries of the triangle formed by force centres Scheme of connection of level controls is shown in fig 8 After placement of the ...

Страница 11: ...ia Phone 370 5 2651474 Fax 370 5 2651483 E mail sales standa lt Manufacturer of Opto Mechanical equipment for research industry and education http www standa lt 11 Fig 9 Installation of level control Fig 10 Installation of level control ...

Страница 12: ...he pipe edges may lead to leaks from the system Insert the pipes in the collet fittings until tight Pull the pipes accurately to make sure that the connection is secure The length of the pipes between the fittings must be optimal with a small extra length Place the tubes along the braces system and secure them with holders of quick installation fig 11 Connect the air supply and set the pressure co...

Страница 13: ...check the distance between the plate and the support see fig 12 Dimension B must be 7 1 5 mm Set height B of every control by means of the control screws When all controls are set check all three dimensions B once again and adjust them if needed Fig 12 Installation of level control Make sure that the table is floating freely on the supports Move the table accurately from side to side by approximat...

Страница 14: ... support may become a zone of danger Any object present in this zone may be squeezed in the case of changes in the load on the table or in the pressure which can result in injuries 3 2 Adjustment of parameters of the system If a load has a high weight centre the table may start to oscillate In this case you need to reduce the pressure in the system and close the needle valve This will improve the ...

Страница 15: ...inimum maintenance As a rule the maintenance includes the assurance of the cleanliness of the air supplied i e replacement of the filter and removal of any water 3 3 1 Cleaning The table and other parts of the system may be cleaned using a non abrasive liquid detergent 3 3 2 Air quality Oil water or impurities contained in the air entering the system may clog level controls and worsen the operatio...