background image

                                                         

 

 13

 

 

Minify the aperture below 60% of the objective’s numerical aperture unless in a special case, for 

instance, observating an almost transparent specimen. 

  The

 numerical aperture is marked on the objective.

 

For example, the mark “ 10/0.25 ” means the 

magnification is 10×

and the numerical aperture  is  0.25.

 

 

If you want to observe the image of the aperture iris diaphragm, remove one eyepiece and look 

through the tube. You will see a dark circle encroaching on the bottom of the tube. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

3.3.6 Adjusting the Interpupillary 
Distance

Fig.22

 

    The interpupillary distance range: 

55mm

75mm

 

While looking through the eyepieces, 

move both sliders to adjust interpupillary 

distance until the left and right fields of 

view coincide completely. The value on 

the right slider is your interpupillary 

distance.

 

3.3.7 Adjusting the Diopter

Fig.23

 

Turn the diopter rings

 on both eyepieces to 

align the number on the rings (which is 

corresponding to    your interpupillary distance) 

with the white dot

 on both sliders to bring 

the specimen into focus. 

   

The diopter range of the eyepiece is

 

 ±5 

diopter.The number aligned to the line 

on the viewing head is the diopter in use.

 

Fig.23 

 

Fig.22 

Содержание STM-2030BD

Страница 1: ...gital biological microscope To insure safety and obtain optimum performance and familiarize yourself fully with the use of this microscope We recommend that you read the manual thoroughly before opera...

Страница 2: ...er Notice 2 1 Components 4 2 Assembly 5 2 1 Assembly Diagram 5 2 2 Assembly Steps 6 3 Adjustment Operation 10 3 1 Adjustment Sets 10 3 2 Operation 11 4 Specification Table 14 5 Outfit 15 6 Troubleshoo...

Страница 3: ...it will bring a serious damage to the instrument 8 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to disassemble them y...

Страница 4: ...re 8 Do not disassemble any parts of the microscope which will affect the function or decline the performance of the microscope 9 Place the instrument in a cool dry position When not using the microsc...

Страница 5: ...4 1 Components STM 2030BD Digital Biological Microscope Eyepiece Digital viewing head Nosepiece Objectives Main body Condenser Collector Stage Ruler...

Страница 6: ...s The number in the figure shows the assembly steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with hexagon wrench provided When replacing th...

Страница 7: ...ieces into the eyepiece tube until they are against each other as shown in Fig 4 Fig 4 Fig 3 NOTE Operation Conditions 1 Temperature 0 40 Maximum Relative Humidity 85 2 High Temperature High Temperatu...

Страница 8: ...Clean the objective regularly for lens is susceptible to dust When operating use 10 magnification objective to search and focus specimen firstly then replace with higher magnification objective if ne...

Страница 9: ...Plug the power cord into the power supply receptacle safely Make sure the connection is well Do use the supplied power cord all the time If lost or damaged select the same standard cord please 2 2 6...

Страница 10: ...halogen Lamp 6V20W Replace the lamp as follows if necessary 1 Please set the main switch to O off state before replacing and make sure the bulb the lamp room and periphery are all cool enough to carry...

Страница 11: ...15 Right Coarse Focus Knob Right Fine Focus Knob Tension Adjustment Collar Fig 14 Condenser Adjustment Knob Longitudinal Adjustment Knob Transversal Adjustment Knob Locking Set Left Fine Focus Knob Le...

Страница 12: ...f you finish the observation with the short working distance objective and want to change another one be careful of not letting the objective touch the specimen 3 3 3 Focusing the specimen 18 19 1 Foc...

Страница 13: ...r has been adjusted too It also needn t any user s operation 3 3 5 Aperture Iris Diaphragm Adjustment Fig 20 21 Turn the aperture iris diaphragm lever to adjust the aperture iris diaphragm Generally s...

Страница 14: ...3 3 6 Adjusting the Interpupillary Distance Fig 22 The interpupillary distance range 55mm 75mm While looking through the eyepieces move both sliders to adjust interpupillary distance until the left an...

Страница 15: ...oving Range 30mm Fine Division 0 004mm Condenser Abbe Condenser NA 1 2 Stage Double Layers Mechanical Stage 140mm 140mm Moving Range 74 50mm Illumination 6V20W Halogen Lamp 4 2 Eyepieces and Objective...

Страница 16: ...nal Condenser Abbe Condenser NA 1 2 1 Optional Nosepiece Quadruple 1 Optional Illumination 6V20W Halogen Lamp 1 Optional Spare lamp 6V20W Halogen lamp 2 Optional Spare fuse 2A 2 Optional Eyepieces 10...

Страница 17: ...oo thick or too thin Use the standard coverslip 0 17mm The specimen is placed inversely Reversal it back There was oil on the dry objective easily happened in 40X objective Clean it up There are stain...

Страница 18: ...er magnification The specimen is placed inversely The coverslip is too thick Turn inversely Use the standard coverslip 0 17 mm The specimen is not easy to move The specimen holder is not fixed Fix it...

Страница 19: ...18 The bulb flickers or the brightness is vertiginous The bulb is going to burn out Replace it The bulb is not entirely inserted into the holder Check and insert it again...

Отзывы: