ST DELTA SOLAR Скачать руководство пользователя страница 15

12

11. Temperature measurement

The

 

DELTA SOLAR

allows you to measure temperature with Pt100 and  Pt1000 

temperature sensors in the range from-200(-100)   C...+850  C.

Set the function selectors witch(6) to "

Ω 

."

Connect the sensor to the two terminals.
Briefly press the yellow multifunction push button(5).
The clamp meter switches to temperature measurement, it
automatically detects the connected sensor(Pt100 to Pt1000) and
shows the measured temperature in

0

C on the digital display.

Notes:

This measurement automatically considers the lead resistance of
temperature sensors which are available as accessory.
It is not possible to switch over to temperature measurement when the
30

Ω 

resistance range is selected.

Sensor lead resistance upto 50

Ω

Lead resistance of sensors having a value differing from tha t of company sensors 
can be considered upto a value of 50

Ω

as follows:

Briefly press the yellow multi-function push button(5) again.
The LCD now displays the resistance value which the clamp meter
automatically considers after selecting the temperature measuring range. 
We can recognise that this is the resistance correction value on the
temperature measuring range. The “

0

C ”  character is simultaneously 

shown on the display.
You can set the lead resistance correction value as follows:
Press the HOLD push button(3) to increment the value, or the AUTO/MAN

push button(4) to decrement the value. Each time the push button is briefly

 pressed, the value changes by one digit.

Briefly press the yellow multi-function push button (5) again.
The LCD displays the measured temperature.The flashing decimal point shows 
you that we have entered a correction value for the lead resistance.
The correction value is retained as long as clamp meter is switched on.

Note:

Each time the yellow multi-function push button(5) is briefly pressed, the 
display changes between measured temperature and correction value of 
the lead resistance.

We can exit the temperature measurement function

by pressing the yellow multi-functions button(5) longer, this is confirmed by the

 two sound signals.

by changing the function selector switch.

For the lead resistance, the actual value measured on the clamp meter should 
be taken as correction value and not any specified value.

12. Capacitance measurement

Verify that the device under test is electrically dead. External voltages would 

falsify the measured results. 

Set the function selector switch (6) to 'F'.

Connect the (discharged!) device under test to the"   "and "F" socket via test lead.

Notes:

Connect polarized capacitors with the"_"pole to the"    "socket.
Resistors and semiconductor  junction in parallel with the capacitor falsify the
measured results!

0

0

Содержание DELTA SOLAR

Страница 1: ...True RMS Digital meter Clamp User Manual DELTA SOLAR...

Страница 2: ......

Страница 3: ...DELTA SOLAR 20 21 ON 22 ON 9 19 17 8 16 18 14 10 15 12 13 11...

Страница 4: ...ange 20 Rotary mechanism for clamp jaws 21 Safe trigger mechanism 22 Limit of safe access for hand held 1 Introduction 2 2 Safety features and safety precautions 2 3 Switching the Clamp meter ON 3 4 F...

Страница 5: ...tage composite voltages may exist there Be particularly careful when measuring on HF circuits Dangerous Measurements under moist environmental conditions are not permitted For safe voltage measurement...

Страница 6: ...ble l when the meter shows obvious signs of damage l when the meter no longer functions correctly l after prolonged storage in adverse conditions l due to severe stress during transportation ble take...

Страница 7: ...the best resolution When switching to The meter switches automatically to the next highe range r at 3099 digits 1digit the next lower range at 240 280 digits 1digit 4 2 Manual range selection You can...

Страница 8: ...ues and for calibration procedures The analog indicator has its own polarity indication When measuring DC quantities the analog scale 15 has a negative range of 4 scale divisions so that variations of...

Страница 9: ...ge facility With the MIN MAX function you can hold the minimum and the maximum measured value which was applied to the input of the Clamp Meter after activating MIN MAX function The most important app...

Страница 10: ...stored MAX value same as1 Cleared MIN and MAX ash MIN MAX same as1 Cleared Meter acknowledgement Display MIN MAX Meas uring ranges Sound Signal 1 Activate and Store 2 Store and display 3 Return to 1...

Страница 11: ...DELTA SOLAR DELTA SOLAR 8 Voltage measurement on electrical systems upto 1000V AC And 1500V DC with the Ks30 measuring adapter AUTO DC V 0 20 30 AUTO DC V 0 20 30 5 10 5 10...

Страница 12: ...th a high voltage probe provided the necessary safety precautions are taken Notes The 30 mV and 300 mV measuring ranges can only be selected manually with the AUTO MAN push button 4 On the 1000 V rang...

Страница 13: ...AN AUTO DC V 0 5 20 30 HOLD ON OFF W 0 C 0 C AUTO MAN 10 AUTO DC V 0 5 0 2 30 0 C 0 C HOLD ON OFF AUTO MAN 10 10 Diode Test Forward direction Reverse direction Continuity Test Resistance measurement A...

Страница 14: ...disappears from the LCD When selecting the function Diode test and Continuty test with the function Selector switch 6 the buzzer is always switched OFF Repeated brief pressing of the multifunction pu...

Страница 15: ...set the lead resistance correction value as follows Press the HOLD push button 3 to increment the value or the AUTO MAN push button 4 to decrement the value Each time the push button is brie y presse...

Страница 16: ...cations for the lowest measurable frequencies and the maximum permissible voltages Changing over between voltage frequency and duty cycle measurement Repeated brief pressing of the yellow multi functi...

Страница 17: ...owledges zero setting by a sound signal the LCD shows 00 00 000 0 1digit The displayed current at the instant the push button is pressed is used as reference value It is automatically deduced from val...

Страница 18: ...ertical bus bars over head cables cables in congested places user connects the clamp meter but the keys and display may not be visible hence user is not able to take the readings or operate the keys T...

Страница 19: ...urements on bare bus bar or bare conductor the user s hand comes very close to bare bus bar conductor which increases the risk of electric shocks to the user Also in conventional clamp meters trigger...

Страница 20: ...500V DC AC eff rms sine wave 0 5 32 0 5 3 0 25 1 0 25 1 0 25 1 0 5 3 0 75 2 10 300 Digit 0 75 1 300 Digit 0 1 0 C 0 1 0 C 0 1 0 C 0 1 0 C pt 100 0 C pt 1000 200 0 200 0 0 C 200 0 850 0 0 C 100 0 200 0...

Страница 21: ...uence quantity Range of In uence Measuring ranges Attenuation Common mode interference voltage Normal mode interference voltage V 120 dB 3V 30V 70 dB 300V 70 dB 1000V 60 dB V 50 dB V 110 dB Noise qua...

Страница 22: ...3 Crest 1 of rdg 3 of rdg 0 C 0 1 x speci ed accuracy 10 30Hz 45Hz 66 Hz 400 Hz 15Hz 30Hz 1 of range 1 1 With Temperature Error for data apply per 10 K change in temperature With Frequency Error data...

Страница 23: ...e dioxide cell according to IEC 6F22 alkaline manganese cell according to IEC6LR61 or suitable NiCd storage battery Lifespan Without backlit ON using alkaline manganese cell approx 220 hours on V appr...

Страница 24: ...er verify that the battery of clamp meter does not leak Repeat this check in regular short intervals If the battery leaks completely remove the battery electrolyte carefully with a moist cloth and ins...

Страница 25: ...s Snap the battery connectors to a new 9 V battery and insert the battery in to the battery compartment Replace the battery cover by tting it in to slots on battery compartment Tighten the battery cov...

Страница 26: ...he warranty is not valid in following cases 1 Warranty card duly signed stamped and original cash memo invoice are not sent along with the meter accessories 2 Complete warranty card is not presented t...

Страница 27: ...rsion No STIUS 2020 01 Sifam Tinsley Instrumentation Inc 3105 Creekside Village Drive Suite No 801 Kennesaw GA 30144 USA Contact No 1 404 736 4903 E mail Id psk sifamtinsley com Web www sifamtinsley c...

Отзывы: