An electrical reset condition may occur at temperatures below -20°C.
After cold storage, allow the device to reach room temperature before charging the capacitors,
programming, or implanting the device because cold temperature may affect initial device function.
Its "use before" date has expired, because this can adversely affect pulse generator longevity or
device sterility.
Do not resterilize the pulse generator using an autoclave, gamma-irradiation, organic cleaning agents
(e.g., alcohol, acetone, etc.), or ultrasonic cleaners.
Sterilization Instructions
Sterilization Instructions
Sterilization Instructions
Sterilization Instructions
Contact St. Jude Medical if resterilization is necessary.
Directions for Use
Directions for Use
Directions for Use
Directions for Use
Pulse generator operating characteristics should be verified at the time of implantation and recorded in
the patient file. Complete the Patient Registration Form and return it to St. Jude Medical as it provides
necessary information for warranty purposes and patient tracking.
Copies of this user's manual can be obtained by contacting your St. Jude Medical representative.
Radiopaque Identification
Radiopaque Identification
Radiopaque Identification
Radiopaque Identification
Each pulse generator has an X-ray absorptive marker for non-invasive identification. The marker consists
of the St. Jude Medical logo (SJM) and a two-letter model code.