Table 4. Possible causes and solutions for suboptimal RF
communication
Possib
Possib
Possib
Possible Causes
le Causes
le Causes
le Causes
Solutions
Solutions
Solutions
Solutions
People or objects interfere with
the communication between the
Merlin Antenna and the device.
Make sure that the space
between the Merlin Antenna
and the device is free from
interfering objects/people.
The Merlin Antenna is too far
away from the device.
Move the Merlin Antenna closer
to the device.
Someone is holding the Merlin
Antenna.
Place the Merlin Antenna on a
flat surface. Do not hold the
Merlin Antenna.
Other products in the vicinity
are causing electromagnetic
interference (EMI).
Power off or remove equipment
that could cause EMI.
The Merlin Antenna cable is
wound around the Merlin
Antenna.
Make sure the Merlin Antenna
cable is not wound around the
Merlin Antenna.