13
Table 6. Guidance and manufacturer's declaration - electromagnetic immunity (conducted RF
and radiated RF)
Test
IEC 60601 Test
Level
2F
3
Immunity
Compliance Level
Electromagnetic Environment -
Guidance
3F
4
Conducted
RF
IEC 61000-
4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
[V
1
=3]
Portable and mobile RF communications
equipment should be used no closer to
any part of the Merlin@home transmitter,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance:
(80 MHz to 800 MHz)
(800 MHz to 2.7 GHz)
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey*, should be less than the
compliance level in each frequency
range.**
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF
IEC 61000-
4-3
3 V/m
80 MHz to
2.7 GHz
3 Vrms
[E
1
=3]
3
At 80 MHz and 800 MHz, the higher frequency range applies.
4
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.