25
For devices without the Low Frequency Attenuation Filter, the default Atrial Sensitivity
setting and the lowest possible setting of Ventricular Sensitivity, 0.2 mV, may be more
susceptible to EMI, according to testing required by CENELEC standard EN45502-2-2.
The devices comply with the electromagnetic compatibility requirements of CENELEC
standard EN45502-2-2 at atrial and ventricular sensitivities of 0.3 mV and less sensitive
settings.
For devices with the Low Frequency Attenuation Filter, the default Atrial Sensitivity
setting, the lowest possible setting of Ventricular Sensitivity, 0.2 mV, and the Ventricular
Sensitivity setting of 0.3 mV when the Low Frequency Attenuation Filter is On, may be
more susceptible to EMI, according to testing required by CENELEC standard EN45502-
2-2. The devices comply with the electromagnetic compatibility requirements of
CENELEC standard EN45502-2-2 at atrial sensitivities of 0.3 mV, ventricular
sensitivities of 0.3 mV (Low Frequency Attenuation Filter OFF) or ventricular sensitivities
of 0.4 mV (Low Frequency Attenuation Filter On), and less sensitive settings.
Physician Training
Physicians should be familiar with sterile pulse generator implant procedure and with follow-up
evaluation and management of patients with an ICD or CRT-D (or should refer the patient to such a
physician).