12
MR Conditional system and according to instructions in the MRI Procedure Information document for
the St. Jude Medical™ MR Conditional system.
Contraindications
Contraindications for use of the pulse generator system include ventricular tachyarrhythmias resulting
from transient or correctable factors such as drug toxicity, electrolyte imbalance, or acute myocardial
infarction.
Warnings and Precautions
Resuscitation Availability.
Do not perform device testing unless an external defibrillator and medical
personnel skilled in cardiopulmonary resuscitation (CPR) are readily available.
Lead system.
Do not use another manufacturer’s lead system without demonstrated compatibility as
undersensing cardiac activity and failure to deliver necessary therapy may result.
Avoiding shock during handling.
Disable tachyarrhythmia therapy (Enable/ Disable Tachy Therapy) or
program tachyarrhythmia therapies Off during surgical implant and explant or post-mortem procedures
as well as when disconnecting leads as the device can deliver a serious shock if you touch the
defibrillation terminals while the device is charged.
Additional pacemaker implanted.
These devices provide bradycardia pacing. If another pacemaker is
used, it should have a bipolar pacing reset mode and be programmed for bipolar pacing to minimize