SSET CE-C Series Скачать руководство пользователя страница 4

 

 

  

http://www.ce-transducer.com

                    3rd Floor, building 1, Zaimao industrial park, Baoji road, Bantian, Longgang district, 

[email protected]

                            Shenzhen City, China      Post code: 518049

 

2.5V, a lagging power factor and capacitive load, C = Vo

÷

2.5V; 

If the output is 0-4mA-20Ma, when output 12mA, a advance 

power factor and inductive load, C = (20 mA -Io)

÷

8 mA; when 

output less than 12 mA, a lagging power factor and capacitive 

load, C = (Io-4 mA)

÷

8 mA. 

6The monitoring meter  measures the DC voltage output value Vo 

or the current output value Io. 

Vo-Vz

│≤

25mV is normal, or excessive (0-5V output, 0.5); 

Io-Iz

│≤

80uA is normal, or excessive (4-20mA output, 0.5); 

Io-Iz

│≤

100uA is normal, or excessive (0-20mA output, 0.5); 

7 Repeat the NO.4 and NO.5 operations, the resulting point value 

Vo-Vz

│≤

25mV  or 

Io-Iz

│≤

80uA  (100uA),  the  accuracy  level 

of transducer is qualified. 

Note: 

please consult with our company for the verification method 

of other technical indicators.   

11 Notes

 

1

 

Please  pay  attention  to  the  power  supply  information  on  the 

product label, and the power supply used grade of the transducer, 

otherwise it will cause the product to be damaged. 

2  Transducer  for  the  integrated  structure,  not  removable,  and 

should avoid collision and fall. 

3

 

The  transducers  are  used  in  environments  with  strong 

electromagnetic  interference.  Standard  precaution  such  as 

shielding  the  input  and  /or  output  lines  should  be  observed.  All 

lines  should be  as  short  as  possible.  If  a  group  of  transducers  are 

mounted together, keep a space more than 10mm between adjacent 

units. 

4 The input value given on the transducer label refers to the RMS 

value of the ac signal. 

5

 

Only  use  the  effective  terminal  of  the  transducer.  The  other 

terminals  may  be  connected  with  the  internal  circuit  of  the 

transducer, and can

t be used for other purposes. 

6  Transducer  has  a  certain  anti-lightning  ability,  but  when  the 

transducer  input  and  output  feeders  exposed  to  extreme  bad 

environments, must be taken lightning protection

 

measures. 

7  Don

t  damage  or  modify  the  product  label  and  logo.  Don

disassemble or modify the transmitter, otherwise the company will 

no  longer  provide  the  product  "three  guarantees"  (replacement, 

return, repair) services. 

8  The  transducers  use  flame-retardant  ABS  plastic  shell  package. 

which limit temperature is +75 

. The shell will be deformed with 

high-temperature  baking,  and  will  affect  product  performance.  Do 

not  use  or  save  the  product  near  the  heat  source.  Do  not  bake  the 

product in a high-temperature oven. 

9  When  measuring  the  voltage  or  current  with  the  multi  meter  pen, 

please  screw  the  terminal  screw  in  the  end,  otherwise  it  may  not 

measure the voltage or current output value. 

 

 

 

 

 

 

Отзывы: