4.

When measurement is complete, turn the power switch off. 

 

Adjustment and Maintenance

 

 

Clean the measurement stand periodically

 

Troubleshooting

 

 
If the device does not operate correctly, it may be the result of one of the following. 

 
■ 

The screen is not displayed after switching the power on. 

Cause: The AC adapter is broken. 
Remedy: Check whether the AC adapter is outputting DC 12 V.

 

 
■ 

An overflow is displayed when the touch button is pressed. 

Cause: The connector is disconnected. 
Remedy: Check whether the connector is disconnected from the measurement stand or touch 
button. 

Directly connect the measurement stand and touch button and check whether a value 

between 0.00 and 0.02 is displayed.

 

 

Specifications

 

 

Weight: main device

 0.9 

kg 

: with supports and 

angle adjustment 

1.2 kg 

: measurement stand 

4.5 kg 

Measurement voltage 

DC 10 V 

Function check 

1 M

 

Measurement range 

0.00 to 200 M

 

Power supply 

DC 12 V AC adapter (AC 100 to 240 V) 

Measurement precision 

±10% + 2 digits 

Maximum setting 

1, 5, 10, 20, 35, 50, 100 (M

Minimum setting 

0.1, 0.5, 1, 5, 10 (M

 

 

 

 

 

 

 

Содержание SHOETESTER II

Страница 1: ...future Functions The SHOETESTER II is a dedicated device for measuring the electrical resistance of antistatic shoes for preventing human body electrostatic charging while they are worn Characteristics 1 The device can be used in various ways such as mounted on a wall placed on a desk or combined with the measurement stand 2 Maximum and minimum values can be set with the device warning the user wi...

Страница 2: ...organic solvents such as thinner Locations where the device may be exposed to corrosive gas Locations subject to flames or explosions Locations subject to frequent vibrations Locations subject to sudden changes in temperature or humidity Locations subject to condensation Locations where the device may be exposed to water or oil Maintenance Clean the touch buttons and measurement stand periodically...

Страница 3: ...nt stand Wall mounting rack Power switch Touch button Self check button Operation setting Power input connector Insert the dedicated AC adapter here Earth terminal Result output terminal Insert the result signal cable here optional Measurement signal connector Insert the dedicated measurement signal cable here RS232C Measurement result display area Minimum setting switch Minimum setting switch ...

Страница 4: ...l plate and then press the touch button Since the resistance of shoes varies according to the temperature humidity and dirt and friction on the bottom of the shoe perform periodic measurements and maintenance management Note that the resistance value will be higher when thick socks are worn The measured value may differ if your body is touching a metal plate or other objects other than the touch b...

Страница 5: ...surement signal connector on the bottom of the main device 2 Connect the AC adapter to the power connector of the main device 3 Connect the earth lead to the earth connector and ground it securely Insert the earth lead to the earth connector securely and connect the clip to a grounding object Make sure to ground the earth terminal to ensure accurate measurements ...

Страница 6: ...included dust protection sticker Setting the Maximum and Minimum Values for Alerts Set the maximum value and minimum value as the standard for determining the measurement result The factory default setting is 0 not set for both switches Minimum setting switch Switch number 0 1 2 3 4 5 6 7 8 9 Setting value MΩ Not set 0 1 0 5 1 5 10 Not set Not set Not set Not set Maximum setting switch Switch numb...

Страница 7: ...nce is measured so the device can self diagnose whether it is working properly The device is working properly if a measurement between 0 95 and 1 05 MΩ is displayed in the measurement result area NOTE Perform the self check operation when the device has stabilized after turning the power on ...

Страница 8: ...result is indicated with a light If the measured value is within range OK light green If the measured value exceeds the maximum value HI light red If the measured value exceeds the minimum value LOW light red NOTE If the device is set to make a sound it will emit an operation complete sound or warning sound Refer to Device Operation Settings for details NOTE Refer to the following when connecting ...

Страница 9: ...ch button is pressed Cause The connector is disconnected Remedy Check whether the connector is disconnected from the measurement stand or touch button Directly connect the measurement stand and touch button and check whether a value between 0 00 and 0 02 is displayed Specifications Weight main device 0 9 kg with supports and angle adjustment 1 2 kg measurement stand 4 5 kg Measurement voltage DC 1...

Страница 10: ...onsibility of SHISHIDO ELECTROSTATIC For any queries relating to the product contact the sales office where you purchased the product SHISHIDO ELECTROSTATIC LTD http www shishido esd co jp HEAD OFFICE MARUNOUCHI BLDG 9F 918 4 1 MARUNOUCHI 2 CHOME CHIYODA KU TOKYO 100 6309 SALES DEPARTMENT TOKYO BRANCH EXPORT DIVISION 3 3 HIGASHI YUKIGAYA 1 CHOME OTA KU TOKYO 145 0065 TEL 81 3 3727 0161 FAX 81 3 37...

Отзывы: