SSD SHOETESTER II Скачать руководство пользователя страница 2

Safety Precautions

 

Make sure to follow the information in this manual for your safety. Areas indicated with a 

 

contain information relating to safety.

 

This device does not conform to explosion-proof specifications. Do not install it 
in locations where flammable gases or solvents are handled, such as painting 
booths etc. Doing so may result in fire or explosion.   

This device is a precision electrical instrument. Avoid installing it in wet, oily, 
hot, and humid locations. In particular, avoid locations of high humidity and 
condensation. There is a possibility of fire due to breakdown.

   

Never use the device if you have a medical device such as a pacemaker in your 

body. The weak electric current that flows into the body from the device may 
damage such medical devices.

 

 

Installation  

 

Do not use this device in the following locations, as doing so may cause malfunctions.

 

Locations subject to high or low 

temperature, or high humidity 

 

Dusty locations 

 

Locations where the device may be 

exposed to organic solvents such as 

thinner 

 

Locations where the device may be 

exposed to corrosive gas 

 

Locations subject to flames or 

explosions 

 

Locations subject to frequent vibrations 

 

Locations subject to sudden changes in 

temperature or humidity 

 

Locations subject to condensation 

 

Locations where the device may be 

exposed to water or oil

Maintenance  

 Clean the touch buttons and measurement stand periodically. Failure to do so may result in 

incorrect measurements. 

  Make sure to turn the main power of the device OFF before cleaning. 

 

Handling  

  Make sure to connect the earth wire to an appropriate place. Accurate measurements are not 

possible if the earth wire is not connected. 

  Do not disassemble or modify the device. 

  The battery discharges a small amount of electricity even when the device is turned off. If you 

do not intend to use this device for an extended length of time, remove the battery. 

  If the device emits any abnormal odors or sounds, smoke, or heat, turn OFF the main power 

immediately, and contact your point of purchase. Failure to do so may result in fire or a short 

circuit. 

  Do not remove name plates or labels. 

  Do not do anything with the device that is not described in this manual.

 

Содержание SHOETESTER II

Страница 1: ...future Functions The SHOETESTER II is a dedicated device for measuring the electrical resistance of antistatic shoes for preventing human body electrostatic charging while they are worn Characteristics 1 The device can be used in various ways such as mounted on a wall placed on a desk or combined with the measurement stand 2 Maximum and minimum values can be set with the device warning the user wi...

Страница 2: ...organic solvents such as thinner Locations where the device may be exposed to corrosive gas Locations subject to flames or explosions Locations subject to frequent vibrations Locations subject to sudden changes in temperature or humidity Locations subject to condensation Locations where the device may be exposed to water or oil Maintenance Clean the touch buttons and measurement stand periodically...

Страница 3: ...nt stand Wall mounting rack Power switch Touch button Self check button Operation setting Power input connector Insert the dedicated AC adapter here Earth terminal Result output terminal Insert the result signal cable here optional Measurement signal connector Insert the dedicated measurement signal cable here RS232C Measurement result display area Minimum setting switch Minimum setting switch ...

Страница 4: ...l plate and then press the touch button Since the resistance of shoes varies according to the temperature humidity and dirt and friction on the bottom of the shoe perform periodic measurements and maintenance management Note that the resistance value will be higher when thick socks are worn The measured value may differ if your body is touching a metal plate or other objects other than the touch b...

Страница 5: ...surement signal connector on the bottom of the main device 2 Connect the AC adapter to the power connector of the main device 3 Connect the earth lead to the earth connector and ground it securely Insert the earth lead to the earth connector securely and connect the clip to a grounding object Make sure to ground the earth terminal to ensure accurate measurements ...

Страница 6: ...included dust protection sticker Setting the Maximum and Minimum Values for Alerts Set the maximum value and minimum value as the standard for determining the measurement result The factory default setting is 0 not set for both switches Minimum setting switch Switch number 0 1 2 3 4 5 6 7 8 9 Setting value MΩ Not set 0 1 0 5 1 5 10 Not set Not set Not set Not set Maximum setting switch Switch numb...

Страница 7: ...nce is measured so the device can self diagnose whether it is working properly The device is working properly if a measurement between 0 95 and 1 05 MΩ is displayed in the measurement result area NOTE Perform the self check operation when the device has stabilized after turning the power on ...

Страница 8: ...result is indicated with a light If the measured value is within range OK light green If the measured value exceeds the maximum value HI light red If the measured value exceeds the minimum value LOW light red NOTE If the device is set to make a sound it will emit an operation complete sound or warning sound Refer to Device Operation Settings for details NOTE Refer to the following when connecting ...

Страница 9: ...ch button is pressed Cause The connector is disconnected Remedy Check whether the connector is disconnected from the measurement stand or touch button Directly connect the measurement stand and touch button and check whether a value between 0 00 and 0 02 is displayed Specifications Weight main device 0 9 kg with supports and angle adjustment 1 2 kg measurement stand 4 5 kg Measurement voltage DC 1...

Страница 10: ...onsibility of SHISHIDO ELECTROSTATIC For any queries relating to the product contact the sales office where you purchased the product SHISHIDO ELECTROSTATIC LTD http www shishido esd co jp HEAD OFFICE MARUNOUCHI BLDG 9F 918 4 1 MARUNOUCHI 2 CHOME CHIYODA KU TOKYO 100 6309 SALES DEPARTMENT TOKYO BRANCH EXPORT DIVISION 3 3 HIGASHI YUKIGAYA 1 CHOME OTA KU TOKYO 145 0065 TEL 81 3 3727 0161 FAX 81 3 37...

Отзывы: