Running the unit from DC power 

1. Disconnect the power cord (1) and press the ON/OFF switch (5) on the control panel to start the vacuum 
pump. 

Adjusting the vacuum level 

• Occlude the patient tube and turn the vacuum regulator (7) clockwise to the stop. 
• Observe gauge (located above the vacuum regulator). The vacuum reading should rise to -300mmHg from 
zero in 3 seconds. It should max out at approximately ≥ -525mmHg. Lower levels of negative pressure will be 
observed at altitude. 

 
Battery Charging from AC Power using AC-DC Charger: 

The optional Fixed Voltage Charger will not run the 

pump. If battery charging from AC power is required, connect the SSCOR AC-DC charger (1) to an AC outlet. 
The charger connects to the unit through the receptacle (1A) on the front panel. 

 
DC Battery Charging from Vehicle 

Connect the DC power cord to the cigarette lighter receptacle in the vehicle or hard wire the DC power cord 
to the DC electrical system of the vehicle on a properly fused line in front of the master switch. Connect the 
electrical line cord to the suction unit by securely attaching the charging plug (1.1A) into the receptacle (1A). 
This charging method is designed to keep the battery charging at all times. If the suction unit is operated 
while it is hooked up to the vehicle it will utilize the vehicle power and save its own battery for emergency 
use. All SSCOR suction units have a diode to prevent drawdown from the pump to the vehicle electrical  
system and a fuse to protect the pump from vehicle electrical surges. If your suction unit is wired to the  
vehicle battery via an automatic load switch power supply, be sure to use a filter in order to eliminate any 
voltage spikes. 

 
Battery Charging Verification 

Once the unit is connected to the charging source, check the control panel (2). The “power source  
connected” LED (4), when lit, shows a good connection and indicates the battery is charging. 

 
Charging the Battery Using the 8310 Charging / Retention Bracket (Model 2309BV Only):  

Hard wire the Model 8310 Charging / Retention Bracket to the hot DC circuit (in front of the master switch). 
Make certain the circuit is properly fused according to appropriate vehicle standards. The vehicle electrical 
system will  furnish power to both run the pump and charge the battery. See page 10 for additional  
information

.

 

SSCOR EVX - Model 2309  

Clearing The Airway Is Our #1 Priority 

Getting Acquainted 

Rev. R (12/13) 

Содержание EVX 2309

Страница 1: ...ng The Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com Email marketing sscor com techsupport sscor com ...

Страница 2: ...ister lid will close down when the canister is filled with fluids 8 Install a new canister before testing for vacuum over 300mmHg to minimize the possibility of implosion which can occur when a canister is aged or damaged 9 SSCOR s battery charger charges batteries only The battery charger will not run the pump 10 External equipment intended for connection to signal input signal output or other co...

Страница 3: ...of power and can be pre set to be activated immediately upon reaching the distressed patient All controls are clearly labeled and easily accessible The EVX is a portable 12V DC battery operated suction device for resuscitation emergencies in prehospital care The battery is recharged by direct connection to the vehicle or an optional SSCOR fixed voltage charger Model 2309 only A fully charged batte...

Страница 4: ...ized Label Attention Consult Accompanying Documents Shipping Carton Manufacturer Shipping Carton Date of Manufacture Shipping Carton Authorized Representative in the European Community Control Panel On Turn Pump On Off Turn Pump Off SSCOR EVX Model 2309 4 Clearing The Airway Is Our 1 Priority Description of Symbols Rev R 12 13 ...

Страница 5: ...1 1A into the receptacle 1A This charging method is designed to keep the battery charging at all times If the suction unit is operated while it is hooked up to the vehicle it will utilize the vehicle power and save its own battery for emergency use All SSCOR suction units have a diode to prevent drawdown from the pump to the vehicle electrical system and a fuse to protect the pump from vehicle ele...

Страница 6: ...attery quality is suspected to ensure proper performance of the device 1 Check the power source connected LED on the control panel 4 is lit and charge the device overnight to ensure the battery is fully charged 2 Remove the power cord and run the unit from its internal DC battery 3 Check for vacuum by occluding the patient tube and set the vacuum regulator 7 to the maximum vacuum setting 4 Observe...

Страница 7: ... REMOVE MAIN CHASSIS SCREWS A 1 screw under the canister holder B 2 screws in the back of the handle B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system H DC receptacle Wiring Harness I Battery Bracket J Battery 12V Sealed Lead Acid K N A L Vacuum pump Return to SSCOR or authorized service center for service Do not ...

Страница 8: ...acuum pump 6 For technical assistance call 800 434 5211 or international 1 818 504 4054 NOTE No part requires lubrication and lubricants should not be used CHARACTERISTICS SPECIFICATIONS Size 17 L x 9 H x 6 W 43 2cm L x 22 9cm H x 15 2cm W Weight 8 6 pounds 3 9 Kgs Vacuum Pump 12V DC 3 0 A Clinical Airflow 30LPM Exceeds 525mmHg Variable Regulator 7 Controls negative pressure from 50 to 525mmHg Gau...

Страница 9: ...istance call 800 434 5211 or 1 818 504 4054 SSCOR EVX Model 2309 9 Clearing The Airway Is Our 1 Priority Maintenance Disinfection As soon as possible after use the single use disposable canister patient tubing and catheter should be discarded according to local regional national requirements for the disposal of hazardous waste materials Clean the exterior of the SSCOR suction unit using a mild det...

Страница 10: ...rging Retention Bracket Put the four 1 1 2 screws through the holes in the charging bracket Model 8310 Reattach the washers and nuts to the screws to secure the Charging Retention bracket in place Base Mount part 8314 7 Mark the hole placement for a base mount and use a 5 16 drill to drill the holes Remove the hardware from the hardware packet Coat the screws with Loctite not supplied and place th...

Отзывы: