Charging/Retention Bracket Mounting Instructions
The Model 8310 Charging/Retention bracket can be mounted by the base or mounted vertically in order to
hold the SSCOR EVX portable suction unit. Mount the retention bracket to a suitable safe mounting surface
so the indicator lights on the control panel are visible to the user. The unit is shipped ready to be mounted
vertically, but if that does not fit your requirements, you can easily modify the unit to mount it by the base.
Note: the mounting bracket will hold the EVX in place to a force of 10G. Select and prepare a mounting
surface, and take care in the mounting procedure, so the bracket and pump will be secure at a force equal
to, or greater than, 10G.
Vertical Mount
Remove the four ¼-20 x 1-1/2” hex head screws from hardware packet. Mark the hole placement for a
vertical mount and use a 5/16”drill to drill the holes. Coat the screws with Loctite (not supplied) before
securing the Charging/Retention Bracket. Put the four 1-1/2” screws through the holes in the charging
bracket (Model 8310). Reattach the washers and nuts to the screws to secure the Charging/Retention
bracket in place.
Base Mount (part #8314-7)
Mark the hole placement for a base mount and use a 5/16” drill to drill the holes. Remove the hardware
from the hardware packet. Coat the screws with Loctite (not supplied) and place the screws through the
bottom of the Bracket, Retention - Counter Mount (#8314-7) and through the holes you have drilled.
Reattach the washers and nuts to secure the Bracket, Retention – Counter Mount (#8314-7) in place.
Mount the charging bracket (Model 8310) to the base mount (#8314-7) using the hardware in the hardware
packet. Coat the screws with Loctite (not supplied) before securing the Charging/Retention Bracket to the
base mount bracket (#8314-7).
Electrical
Hard wire the Charging / Retention Bracket to the DC Electrical System of the vehicle in front of the master
switch. Make certain that the circuit is properly fused according to appropriate vehicle standards.