1. SSCOR suc on units are not designed or intended for use in extended procedures that require prolonged
high vacuum/low air
fl
ow applica ons, as is the case in wound drainage or endoscopic use or in any other
procedure that produces high vacuum levels within an occluded system for an extended period of me.
Turn the suc on unit o
ff
when it is not in use.
2. Federal law restricts this device to sale, distribu on, and use by, or on the order of a physician, emergency
medical technician, or other medical prac
oner. For use by medical personnel trained in suc oning
techniques and in the use of medical suc on equipment.
3. Operator should be thoroughly familiar with these opera ng instruc ons before this device is used.
4. Do not use in the presence of
fl
ammable agents or anesthe cs.
5. Before tes ng for vacuum over
‐
300mmHg look for an expira on date on the canister (where applicable)
and change the canister if the canister has passed the expira on date to minimize the possibility of
implosion, which can occur when a canister is aged or damaged.
6. External equipment intended for connec on to signal input, signal output or other connectors, shall
comply with relevant IEC standard (e.g. IEC 60950 for IT equipment and the IEC 60601 series for medical
electrical equipment). In addi on, all such combina ons–systems – shall comply with the standard IEC
60601
‐
1
‐
1, safety requirements for medical electrical systems. Equipment not complying with IEC 60601
shall be kept outside the pa ent environment, as de
fi
ned in the standard. Any person who connects
external equipment to signal input, signal output or other connectors has formed a system and is there
fore responsible for the system to comply with the requirements of IEC 60601
‐
1
‐
1. If in doubt, contact
quali
fi
ed technician or your local representa ve.
7. A blinking LED on the control panel indicates low ba ery capacity. This means the ba ery has been
subjected to irreparable damage. REPLACE THE BATTERY! Replacement ba ery may be purchased at
shop.sscor.com.
8. Where the integrity of the external protec ve earth conductor arrangement (ground) is in doubt, the unit
shall be operated from its internal electrical power source (ba ery only). Grounding reliability can only be
achieved when connected to an equivalent receptacle marked “Hospital Only” or “Hospital Grade”
SSCOR Duet®
Model 2314 Series
‐
AC suc on unit with ba ery back
‐
up
SSCOR DUET - Model 2314 Series
Rev W (02/17)
2
Clearing The Airway Is Our #1 Priority
Caution-Notice
Revision W (02/17)