4. Check the air
fl
ow
A. Connect tubing from the pa ent port of the canister to a calibrated air
fl
ow meter. Ensure the regulator
gauge is adjusted to the high se ng (fully turned clockwise) and turn the device on. The reading on the
calibrated
air
fl
ow meter should be above 30 LPM.
B. If the device does not produce 30 LPM check to ensure all tubing connec ons are ght, the canister lid is
securely fastened to the canister and all the port covers on the canister lid are secure.
C. If your device does not produce at least 30 LPM air
fl
ow check the air
fl
ow before the canister, at the union
barb
fi
ng at the top of the device (3). The air
fl
ow at the union barb
fi
ng should be at least 48 LPM.
D. If the tubing from the device to the canister is clamped to the device (4), remove the tubing from the
pa ent port of the canister and take the reading at the distal end of the tubing (5). The air
fl
ow at the
distal end of the tubing should be at least 48 LPM.
E. If the device s ll does not meet the air
fl
ow speci
fi
ca on, call SSCOR technical support (1
‐
800
‐
434
‐
5211 or
1
‐
818
‐
504
‐
4054) to trouble shoot.
A er these tests are complete, plug the device in to charge before releasing the device back to the
fl
oor. The
ba ery condi on LED on the control panel should light green when the device is plugged in and the ba ery is
fully charged.
If your device fails to meet any of these speci
fi
ca ons or if you have any ques ons, call SSCOR technical
support at 1
‐
800
‐
434
‐
5211 or 1
‐
818
‐
504
‐
4054.
12
Clearing The Airway Is Our #1 Priority
Preventive Maintenance
SSCOR DUET - Model 2314 Series
Rev W (02/17)