4. Check the air

fl

ow 

A. Connect tubing from the pa ent port of the canister to a calibrated air

fl

ow meter. Ensure the regulator 

 

gauge is adjusted to the high se ng (fully turned clockwise) and turn the device on.  The reading on the 

 calibrated 

air

fl

ow meter should be above 30 LPM. 

B. If the device does not produce 30 LPM check to ensure all tubing connec ons are  ght, the canister lid is 
 

securely fastened to the canister and all the port covers on the canister lid are secure. 

C. If your device does not produce at least 30 LPM air

fl

ow check the air

fl

ow before the canister, at the union 

 barb 

fi

ng at the top of the device (3). The air

fl

ow at the union barb 

fi

ng should be at least 48 LPM.   

 
 
 
 
 
 
 
 
 
D. If the tubing from the device to the canister is clamped to the device (4), remove the tubing from the  
 pa ent port of the canister and take the reading at the distal end of the tubing (5). The air

fl

ow at the  

 

distal end of the tubing should be at least 48 LPM. 

 
 
 
 
 
 
 
 
 
 
 
E. If the device s ll does not meet the air

fl

ow speci

fi

ca on, call SSCOR technical support (1

800

434

5211 or 

 

1

818

504

4054) to trouble shoot. 

 
A er these tests are complete, plug the device in to charge before releasing the device back to the 

fl

oor. The 

ba ery condi on LED on the control panel should light green when the device is plugged in and the ba ery is 
fully charged. 
 
If your device fails to meet any of these speci

fi

ca ons or if you have any ques ons, call SSCOR technical  

support at 1

800

434

5211 or 1

818

504

4054. 

12 

Clearing The Airway Is Our #1 Priority 

Preventive Maintenance 

SSCOR DUET - Model 2314 Series 

Rev W (02/17)  

Содержание 2314 series

Страница 1: ...e Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com shop sscor com Email marketing sscor com techsupport sscor com ...

Страница 2: ...all comply with relevant IEC standard e g IEC 60950 for IT equipment and the IEC 60601 series for medical electrical equipment In addi on all such combina ons systems shall comply with the standard IEC 60601 1 1 safety requirements for medical electrical systems Equipment not complying with IEC 60601 shall be kept outside the pa ent environment as defined in the standard Any person who connects ex...

Страница 3: ...R Duet is a portable 100V 240V AC constant suc on device with a 12V DC ba ery back up The ba ery is charged by an internal DC dual mode ba ery charger The charger monitors the ba ery charges the ba ery only when necessary shuts down the unit if the ba ery is low and signals ba ery condi on A fully charged ba ery at full capacity will power the unit for 45 minutes 10 SSCOR warrants that each new pr...

Страница 4: ...ping Carton Manufacturer Shipping Carton Date of Manufacture Shipping Carton Authorized Representa ve in the European Community Control Panel Power Source Connected Control Panel Pump is On Control Panel Ba ery Level Of Charge Green LED Full Green Yellow LED High Yellow LED Half Yellow Red LED Low Red LED Ba ery Depleted Put device back on charge Control Panel Push On Push Off 4 Clearing The Airway...

Страница 5: ...gauge located above the vacuum regulator The vacuum reading should rise to 300mmHg from zero in 3 seconds It should max out at approximately 525mmHg Lower levels of nega ve pressure will be observed at al tude 3 Adjust the vacuum reading to your desired level If the vacuum does not meet or exceed 525mmHg check for a leak in the system i e tubing barb connec ons or loose canister lid Refer internal...

Страница 6: ...he desired se ng turn the device off and put it back on charge SSCOR recommends replacement of the ba ery a er 3 years Replacement ba ery may be purchased at shop sscor com PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Power Source Connected LED not lit Fuse blown Molex connec ons disconnected Power cord disconnected Replace fuse in power supply or input receptacle Reconnect Molex Reconnect cord Does no...

Страница 7: ...a ery DC Power Medical Switcher AC Power Rechargeable Sealed Lead Acid SSCOR part 80635 100V 240VAC 47 63Hz Uses 1ea 3A 250V fast ac ng fuse Switch 5 Off Standby On membrane switch AC receptacle 1A Uses 2 2 5A 250V fast ac ng fuses Collec on Canister 13 1200cc ml SSCOR part 48041 Pa ent tubing Vinyl tubing 9 32 ID x 72 L 7 1mm ID x 182 88 cm L SSCOR part 43200 Suc on Tip HI D Big S ck SSCOR part 44...

Страница 8: ...essary to open the exhaust door unless fluids have entered the system E PC Board Electrical Circuits Internal Dual Mode Ba ery Charger Do not a empt to service the PC Board Return to SSCOR or authorized service center for service F Power Supply Do not a empt to service the Power Supply Return to SSCOR or authorized service center for service G Func onal ground H AC receptacle Two 250V 2 5A fuses a...

Страница 9: ...xhaust filter is soiled remove the two screws holding the filter door in place see page 8 remove the soiled filter and replace with new filter SSCOR part number 90160 Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunc on Disinfec on Use personal protec ve equipment such as gloves a smock and face and e...

Страница 10: ...ll be secure at a force equal to or greater than 10G Ver cal Mount Remove the four 20 x 1 1 2 hex head screws from hardware packet Mark the hole placement for a ver cal mount and use a 5 16 drill to drill the holes Coat the screws with Loc te not supplied before securing the Charging Reten on Bracket Put the four 1 1 2 screws through the holes in the charging bracket Model 8323 Rea ach the washers...

Страница 11: ...D At the end of the run test turn the device off and reconnect the device to the charging source For a further explana on of the SSCOR ba ery run test visit the SSCOR Website at h p www sscor com training_videos html 3 Determine if the gauge is properly calibrated A er the ba ery has been fully charged when the ba ery condi on LED on the control panel is green connect tubing from the pa ent port of...

Страница 12: ... D If the tubing from the device to the canister is clamped to the device 4 remove the tubing from the pa ent port of the canister and take the reading at the distal end of the tubing 5 The airflow at the distal end of the tubing should be at least 48 LPM E If the device s ll does not meet the airflow specifica on call SSCOR technical support 1 800 434 5211 or 1 818 504 4054 to trouble shoot A er ...

Отзывы: