REMOTE PROGRAMMING
4-4
RNGH(?) {i}
The RNGH command disables (i=0) and enables (i=1) range hold. When
range hold is disabled the SR715/720 will autorange. The RNGH? query
returns the range hold status.
VOLT(?) {x}
The VOLT command sets the drive voltage to x volts. x may range from 0.1 V
to 1.0 V and is rounded to the nearest 0.05 V. The VOLT? query returns the
drive voltage.
MEASUREMENT CONTROL COMMANDS
PREL(?) {x}
The PREL command sets the relative parameter value for deviation and
percent deviation measurements to x. This command will return an error if
the SR715/720 is set to AUTO parameter mode. The units of x are Ohms in
R+Q mode, Henrys in L+Q mode, and Farads in C+D and C+R modes. The
PREL? query returns the present nominal value.
STRT
The STRT command starts a measurement. If a measurement is already in
progress, this command is ignored.
STOP
The STOP command stops the current measurement.
*
TRG
The
*
TRG IEEE common command functions the same as the STRT
command.
MEASUREMENT RESULT COMMANDS
The following commands return the results of the SR715/720's measurements. The results may be returned in
either ASCII or binary format, and may or may not return measurement status information.
OUTF(?) {i}
Set OUTF command sets the format for the X commands to Verbose ASCII
(i=0), Concise ASCII (i=1), Verbose Binary (i=2), or Concise Binary (i=3). The
verbose formats return both measurement value and status information, while
the concise formats return only measurement values. The formats are
detailed below. The OUTF? query returns the present output format.
Verbose ASCII
This format returns both status and measurement data as a string of ASCII
characters. The byte ordering and their meanings are given below.
Byte
Character
Meaning
1
status
The first byte is the measurement status.The first
character can be any of the following:
G
Measurement good. No errors.
I
Invalid measurement, no measurement completed,
A/D converter errors.
L
Measurement overloaded.
U
Underrange. The measurement value is below the
nominal range of values for this measurement range.
O
Overrange. The measurement value is above the
nominal range of values for this measurement range.
R
Out of Range. The value is above the SR715/720's
ability to measure on this measurement range.