ACCURACY
3-2
IMPEDANCE ACCURACY
Accuracy of Z = ± [A x Ki x Kv + (Kh + Kl) x 100] %
A
= Basic Impedance Accuracy Factor
from graph on next page.
If the unit is in the constant voltage
(CV) mode, double the basic
accuracy factor.
Ki = Integration Time Factor (see Ki
Table below)
Kv
=
Drive Voltage Error Factor
(see Kv Table below)
Kh, Kl
=
Extreme Range Error Term (see
Kh,Kl Table below)
Ki Table
Meas Rate Frequency Zm
Ki
slow,med all
1
fast
100 Hz-1 kHz 6.25
Ω
<Zm<400 k
Ω
3
fast
all other
2
Kv Table
Vout (V rms)
Kv
1.0 - 0.55
1 / Vout
0.5 - 0.3
0.5 / Vout
0.25 - 0.15
0.25 / Vout
0.10
0.11/Vout
Note that Kv equals one for the primary drive
voltages (1.0, 0.5 and 0.25 Vrms).
Kh & Kl Table
Frequency
Kl
Kh
100,120,1k Hz (1 m
Ω
/ Zm)
(Zm / 2 G
Ω
)
10 kHz
(1 m
Ω
/ Zm)
(Zm / 1.5 G
Ω
)
100 kHz
(4 m
Ω
/ Zm)
(Zm / 50 M
Ω
)
Zm = the impedance of the device being
measured.
Note that Kl is negligible for impedances above
100
Ω
and Kh is negligible for impedances below
1 k
Ω
, both at all frequencies.