CALIBRATION CONT’D 

 

STEP 5

:  Place a known calibrated weight, traceable to NIST 

standards, onto the weighing surface and compare it to the displayed 
reading.  

Note:

  For best results use a minimum of 10 pounds to 

calibrate.  (DO NOT USE barbell weights or calibrate to a mechanical 
scale.)   

STEP 6

:  Use the right hand CAL button to make “

+

” corrections 

and the left hand calibration button to make “

-

 

corrections to the 

displayed weight.  The displayed value should be within 0.1% of 
the calibrated weight per the tolerance table. 

STEP 7

:  When settings are completed:  Press the 

“HOLD”

 button to SAVE your settings or 

press the “

READ

” button to CANCEL.  Both choices will EXIT CAL

 

mode

STEP 8

:  Re-attach plugs on the bottom of the scale. 

 

 
 
 
 
 
 
 
 
 
 
 

TROUBLESHOOTING 

 
 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Low readings 

Check offset and verify calibration 

Only half of display activates 

Press the “WEIGH” button   

No power 

Check that batteries are making contact with each 
other and the terminals of the battery holder.   

For additional information or assistance, telephone the Service Hotline: 1-800-654-6360 

or e-mail: 

[email protected]

 

 

CALIBRATION 

TOLERANCE TABLE 

LOW 

LIMIT 

APPLIED 

LOAD 

HIGH 

LIMIT 

    .99 

  1.0 

   1.01

  4.99 

  5.0 

   5.01

  9.99 

10.0 

 10.01

14.98 15.0  

15.02

17.98 18.0  

18.02

CAUTION 

The integrated circuits and semiconductors on 
the printed circuit boards may be damaged 
by electrostatic discharge (ESD).  Be sure to 
use proper handling precautions at all times.  

Содержание SR615

Страница 1: ...0210 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SInstruments Inc by SInstrument...

Страница 2: ...LY 4 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 8 BATT...

Страница 3: ...Scale DESCRIPTION QUANTITY SCALE BASE 1 ea SCALE BASE COVER 1 ea CLEAR ACRYLIC CRADLE One cradle size included with scale base 1 ea One of the following choices included CRADLE CLEAR ACRYLIC OPEN END...

Страница 4: ...the six 6 D cell batteries as described on battery compartment cover label and replace the Battery Compartment Cover Turn scale base right side up and place on a stable flat surface so that all four f...

Страница 5: ...sentation of the infant s weight regardless of weight distribution The infant s weight is displayed on a 16 character dot matrix LCD With a push of a button weight data may be viewed in either pounds...

Страница 6: ...cratched by abrasive cleaners Mild soap and water is recommended for general cleaning and disinfecting STORAGE and TRANSPORTATION If storing this equipment for periods longer than three 3 months remov...

Страница 7: ...on AUTO POWER DOWN Approximately 120 seconds HOLD While active freezes display data and stores it temporarily in memory LAST WEIGHT RECALL Press HOLD button to recall last stored displayed reading AVE...

Страница 8: ...t reading after the system has shut down press the HOLD button before pressing any other button The ZERO button will clear the held memory The MODE button allows weight data to be displayed in pounds...

Страница 9: ...displays the patient s weight Zero will clear locked weight The minimum weight for the AUTO HOLD mode to function is 1 10 pounds To enable AUTO HOLD mode press and hold the HOLD button for approximat...

Страница 10: ...og to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the refer...

Страница 11: ...hold both CAL buttons 1 simultaneously for 10 seconds The display will read HOLD TO CAL as the right hand digit counts down to enter the CAL mode STEP 3 Attach cradle Offset figure should read betwee...

Страница 12: ...librated weight per the tolerance table STEP 7 When settings are completed Press the HOLD button to SAVE your settings or press the READ button to CANCEL Both choices will EXIT CAL mode STEP 8 Re atta...

Страница 13: ...s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc...

Страница 14: ...Part No MAN615_110210 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SInstruments...

Отзывы: